Treatment robustness of total body irradiation with volumetric modulated arc therapy.
Seravalli, Enrica; Willemsen-Bosman, Mirjam; Zoetelief, Annelies; Roosenboom, Sanne; Harderwijk, Tessa; Krikke, Lean; Bol, Gijsbert; Kotte, Alexis; Huijboom, Eline; van Loon, Karel; Hoeben, Bianca.
Affiliation
  • Seravalli E; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Willemsen-Bosman M; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Zoetelief A; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Roosenboom S; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Harderwijk T; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Krikke L; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Bol G; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Kotte A; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Huijboom E; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • van Loon K; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
  • Hoeben B; University Medical Center Utrecht, Department of Radiation Oncology, Utrecht, The Netherlands.
Phys Imaging Radiat Oncol ; 29: 100537, 2024 Jan.
Article in En | MEDLINE | ID: mdl-38292651
ABSTRACT
This study evaluated the robustness of multi-isocenter Volumetric Modulated Arc Therapy Total Body Irradiation dose distribution in the overlapping region between the head-first and feet-first computed tomography scans, considering the longitudinal isocenter shifts recorded during treatment delivery. For 15 out of 22 patients, the dose distribution in the overlapping region fulfilled all three the robustness criteria. The overlapping region dose distribution of the remaining 7 cases fulfilled two robustness criteria. The dose distribution was found to be robust against daily recorded longitudinal isocenter shifts, as a consequence of the patient position verification procedure, of up to 16 mm.