Reliability and validity of the MX3 portable sweat sodium analyser during exercise in warm conditions.
Brown, Harry A; Clark, Brad; Périard, Julien D.
Affiliation
  • Brown HA; Research Institute for Sport and Exercise Science (UCRISE), University of Canberra, Bruce, Canberra, ACT, Australia.
  • Clark B; Research Institute for Sport and Exercise Science (UCRISE), University of Canberra, Bruce, Canberra, ACT, Australia.
  • Périard JD; Research Institute for Sport and Exercise Science (UCRISE), University of Canberra, Bruce, Canberra, ACT, Australia. Julien.Periard@canberra.edu.au.
Eur J Appl Physiol ; 124(7): 2153-2160, 2024 Jul.
Article in En | MEDLINE | ID: mdl-38430262
ABSTRACT

PURPOSE:

Accurately measuring sweat sodium concentration ([Na+]) in the field is advantageous for coaches, scientists, and dieticians looking to tailor hydration strategies. The MX3 hydration testing system is a new portable analyser that uses pre-calibrated biosensors to measure sweat [Na+]. This study aimed to assess the validity and reliability of the MX3 hydration testing system.

METHODS:

Thirty-one (11 females) recreationally active participants completed one experimental trial. During this trial, participants exercised at a self-selected pace for 45 min in a warm environment (31.5 ± 0.8 °C, 63.2 ± 1.3% relative humidity). Sweat samples were collected from three measurement sites using absorbent patches. The samples were then analysed for sweat [Na+] using both the MX3 hydration testing system and the Horiba LAQUAtwin-NA-11. The reliability of the MX3 hydration testing system was determined following two measurements of the same sweat sample.

RESULTS:

The mean difference between measurements was 0.1 mmoL·L-1 (95% limits of agreement (LoA) - 9.2, 9.4). The analyser demonstrated a coefficient of variation (CV) of 5.6% and the standard error of measurement was 3.3 mmoL·L-1. When compared to the Horiba LAQUAtwin-NA-11, there was a mean difference of - 1.7 mmoL·L-1 (95% LoA - 0.25 X ¯ , 0.25 X ¯ ) and the CV was 9.8%.

CONCLUSION:

The MX3 hydration testing system demonstrated very good single-trial reliability, moderate agreement and a very good CV relative to the Horiba LAQUAtwin-Na-11. To further validate its performance, the MX3 hydration testing system should be compared with analytical techniques known for superior reliability and validity.