Calibration of three-dimensional ultrasound images for image-guided radiation therapy.
Bouchet, L G; Meeks, S L; Goodchild, G; Bova, F J; Buatti, J M; Friedman, W A.
Affiliation
  • Bouchet LG; Department of Neurological Surgery, University of Florida, Gainesville, USA. bouchet@neurosurgery.ufl.edu
Phys Med Biol ; 46(2): 559-77, 2001 Feb.
Article de En | MEDLINE | ID: mdl-11229734
ABSTRACT
A new technique of patient positioning for radiotherapy/radiosurgery of extracranial tumours using three-dimensional (3D) ultrasound images has been developed. The ultrasound probe position is tracked within the treatment room via infrared light emitting diodes (IRLEDs) attached to the probe. In order to retrieve the corresponding room position of the ultrasound image, we developed an initial ultrasound probe calibration technique for both 2D and 3D ultrasound systems. This technique is based on knowledge of points in both room and image coordinates. We first tested the performance of three algorithms in retrieving geometrical transformations using synthetic data with different noise levels. Closed form solution algorithms (singular value decomposition and Horn's quaternion algorithms) were shown to outperform the Hooke and Jeeves iterative algorithm in both speed and accuracy. Furthermore, these simulations show that for a random noise level of 2.5, 5, 7.5 and 10 mm, the number of points required for a transformation accuracy better than 1 mm is 25, 100, 200 and 500 points respectively. Finally, we verified the tracking accuracy of this system using a specially designed ultrasound phantom. Since ultrasound images have a high noise level, we designed an ultrasound phantom that provides a large number of points for the calibration. This tissue equivalent phantom is made of nylon wires, and its room position is optically tracked using IRLEDs. By obtaining multiple images through the nylon wires, the calibration technique uses an average of 300 points for 3D ultrasound volumes and 200 for 2D ultrasound images, and its stability is very good for both rotation (standard deviation 0.4 degrees) and translation (standard deviation 0.3 mm) transformations. After this initial calibration procedure, the position of any voxel in the ultrasound image volume can be determined in world space, thereby allowing real-time image guidance of therapeutic procedures. Finally, the overall tracking accuracy of our 3D ultrasound image-guided positioning system was measured to be on average 0.2 mm, 0.9 mm and 0.6 mm for the AP, lateral and axial directions respectively.
Sujet(s)
Recherche sur Google
Collection: 01-internacional Base de données: MEDLINE Sujet principal: Planification de radiothérapie assistée par ordinateur / Échographie Type d'étude: Diagnostic_studies / Guideline / Prognostic_studies Limites: Humans Langue: En Journal: Phys Med Biol Année: 2001 Type de document: Article Pays d'affiliation: États-Unis d'Amérique
Recherche sur Google
Collection: 01-internacional Base de données: MEDLINE Sujet principal: Planification de radiothérapie assistée par ordinateur / Échographie Type d'étude: Diagnostic_studies / Guideline / Prognostic_studies Limites: Humans Langue: En Journal: Phys Med Biol Année: 2001 Type de document: Article Pays d'affiliation: États-Unis d'Amérique