Randomized evaluation of the TriActiv balloon-protection flush and extraction system for the treatment of saphenous vein graft disease.
J Am Coll Cardiol
; 46(9): 1677-83, 2005 Nov 01.
Article
de En
| MEDLINE
| ID: mdl-16256868
OBJECTIVES: The Protection During Saphenous Vein Graft Intervention to Prevent Distal Embolization (PRIDE) study compared outcomes with the TriActiv System (Kensey Nash Corp., Exton, Pennsylvania), a balloon-protection flush and extraction device, with an embolic protection group during treatment of saphenous venous grafts (SVGs). BACKGROUND: Treatment of SVGs with embolic protection reduces adverse cardiac events. METHODS: We conducted a prospective trial randomizing 631 patients with coronary ischemia and lesions in SVGs to embolic protection with the TriActiv System or control group (Guardwire System [Medtronic AVE, Santa Rosa, California] or Filterwire EX [Boston Scientific Corp., Maple Grove, Minnesota]). RESULTS: The incidence of major adverse cardiac events at 30 days was 11.2% for the TriActiv group and 10.1% for the control group (relative risk = 1.1%; 95% confidence interval 0.67 to 1.76; p = 0.65; p = 0.02 for non-inferiority). Safety and efficacy end points were similar between groups except that patients randomized to the TriActiv System had more hemorrhagic complications (10.9% vs. 5.4%; p = 0.01). CONCLUSIONS: The TriActiv System was not inferior to approved embolic protection devices for the treatment of diseased SVGs.
Recherche sur Google
Collection:
01-internacional
Base de données:
MEDLINE
Sujet principal:
Complications postopératoires
/
Veine saphène
/
Occlusion par ballonnet
/
Occlusion du greffon vasculaire
Type d'étude:
Clinical_trials
/
Etiology_studies
/
Observational_studies
/
Risk_factors_studies
Limites:
Adult
/
Aged
/
Aged80
/
Female
/
Humans
/
Male
/
Middle aged
Langue:
En
Journal:
J Am Coll Cardiol
Année:
2005
Type de document:
Article
Pays d'affiliation:
États-Unis d'Amérique
Pays de publication:
États-Unis d'Amérique