Use of kilovoltage X-ray volume imaging in patient dose calculation for head-and-neck and partial brain radiation therapy.
Hu, Weigang; Ye, Jinsong; Wang, Jiazhou; Ma, Xuejun; Zhang, Zhen.
Affiliation
  • Hu W; Department of Radiation Oncology, Cancer Hospital, Department of Oncology, Shanghai Medical college, Fudan University, Shanghai, China. zhenzhang6@yahoo.com
Radiat Oncol ; 5: 29, 2010 Apr 19.
Article de En | MEDLINE | ID: mdl-20403191
BACKGROUND: To evaluate the accuracy of using kilovoltage x-ray cone-beam computed tomography (kV-CBCT) imaging for in vivo dose calculations. METHODS: A Region-of-Interest (ROI) CT number mapping method was developed to generate the cone-beam CT number vs. relative electron density calibration curve for 3D dose calculations. The stability of the results was validated for three consecutive months. The method was evaluated on three brain tumors and three head-and-neck tumor cases. For each patient, kV-CBCT images were acquired on the first treatment day and two-week intervals on the Elekta XVI system. The delivered dose distributions were calculated by applying the patients' treatment plans to the kV-CBCT images. The resulting dose distributions and dose volume histograms (DVHs) of the tumor and critical structures were compared to the original treatment plan. RESULTS: The kV-CBCT electron density calibration was stable within 1.5% over a three-month period. The DVH and dose distribution comparison based on the planning CT and the initial kV-CBCT showed good agreements for majority of cases. The doses calculated from the planning CT and kV-CBCT were compared on planes perpendicular to the beam axes and passing through the isocenter. Using gamma analysis with a criterion of 2 mm/2% and a threshold of 10%, more than 99.5% of the points on the iso-planes exhibited gamma <1. For one patient, kV-CBCT images detected 5.8% dose variation in the right parotid due to tumor shrinkage and patient weight loss. CONCLUSIONS: ROI mapping method is an effective method for the creation of kV-CBCT electron density calibration curves for head-and-neck and brain tumor patients. Dose variations as monitored using kV-CBCT imaging suggest that some patients can benefit from adaptive treatment plan re-optimization.
Sujet(s)

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Tumeurs du cerveau / Tomodensitométrie / Tumeurs de la tête et du cou Type d'étude: Observational_studies / Prognostic_studies Limites: Humans Langue: En Journal: Radiat Oncol Sujet du journal: NEOPLASIAS / RADIOTERAPIA Année: 2010 Type de document: Article Pays d'affiliation: Chine Pays de publication: Royaume-Uni

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Tumeurs du cerveau / Tomodensitométrie / Tumeurs de la tête et du cou Type d'étude: Observational_studies / Prognostic_studies Limites: Humans Langue: En Journal: Radiat Oncol Sujet du journal: NEOPLASIAS / RADIOTERAPIA Année: 2010 Type de document: Article Pays d'affiliation: Chine Pays de publication: Royaume-Uni