This paper presents a contactless method for monitoring infant sleep apnoea. The method uses a remote infrared (IR) sensor to monitor the motion of the infant's abdomen. This method has important clinical advantages in comparison with conventional methods. First, it improves the comfort and compliance of the infants. Second, it eliminates the effects of motion artefacts and skin irritation. Third, it enhances infant safety. Fourth, it does not require frequent calibration and thus enables a continuous monitoring of sleep apnoea. Finally, it is suitable for home applications. Experimental evaluation of this method showed that it has 85% accuracy, 85.71% specificity and 84.62% sensitivity, which imply that it is a promising technique for the detection of infant sleep apnoea.
Sujet(s)

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Syndromes d'apnées du sommeil / Traitement du signal assisté par ordinateur / Monitorage physiologique Type d'étude: Diagnostic_studies / Prognostic_studies Limites: Humans / Infant Langue: En Journal: J Med Eng Technol Année: 2010 Type de document: Article Pays d'affiliation: Jordanie Pays de publication: Royaume-Uni

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Syndromes d'apnées du sommeil / Traitement du signal assisté par ordinateur / Monitorage physiologique Type d'étude: Diagnostic_studies / Prognostic_studies Limites: Humans / Infant Langue: En Journal: J Med Eng Technol Année: 2010 Type de document: Article Pays d'affiliation: Jordanie Pays de publication: Royaume-Uni