Your browser doesn't support javascript.
loading
Commissioning of electron cyclotron emission imaging instrument on the DIII-D tokamak and first data.
Tobias, B; Domier, C W; Liang, T; Kong, X; Yu, L; Yun, G S; Park, H K; Classen, I G J; Boom, J E; Donné, A J H; Munsat, T; Nazikian, R; Van Zeeland, M; Boivin, R L; Luhmann, N C.
Affiliation
  • Tobias B; University of California at Davis, Davis, California 95616, USA. bjtobias@ucdavis.edu
Rev Sci Instrum ; 81(10): 10D928, 2010 Oct.
Article de En | MEDLINE | ID: mdl-21033956
ABSTRACT
A new electron cyclotron emission imaging diagnostic has been commissioned on the DIII-D tokamak. Dual detector arrays provide simultaneous two-dimensional images of T(e) fluctuations over radially distinct and reconfigurable regions, each with both vertical and radial zoom capability. A total of 320 (20 vertical×16 radial) channels are available. First data from this diagnostic demonstrate the acquisition of coherent electron temperature fluctuations as low as 0.1% with excellent clarity and spatial resolution. Details of the diagnostic features and capabilities are presented.

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Langue: En Journal: Rev Sci Instrum Année: 2010 Type de document: Article Pays d'affiliation: États-Unis d'Amérique

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Langue: En Journal: Rev Sci Instrum Année: 2010 Type de document: Article Pays d'affiliation: États-Unis d'Amérique