Your browser doesn't support javascript.
loading
Severe complications and failures of incontinence surgery using the Remeex(®) adjustable tension sling (external mechanical regulator).
Lorenzo-Gómez, M F; Padilla-Fernández, B; Virseda-Rodríguez, A J; Collazos-Robles, R E; García-Cenador, M B; Mirón-Canelo, J A.
Affiliation
  • Lorenzo-Gómez MF; Servicio de Urología, Complejo Asistencial Universitario de Salamanca, Salamaca, España; Departamento de Cirugía, Universidad de Salamanca, Salamanca, España; Instituto de Investigación Biomédica de Salamanca (IBSAL), Salamanca, España. Electronic address: mflorenzogo@yahoo.es.
  • Padilla-Fernández B; Servicio de Urología, Complejo Hospitalario Universitario de Canarias, Tenerife, España.
  • Virseda-Rodríguez AJ; Servicio de Urología, Complejo Asistencial Universitario de Salamanca, Salamaca, España; Instituto de Investigación Biomédica de Salamanca (IBSAL), Salamanca, España.
  • Collazos-Robles RE; Servicio de Ginecología y Obstetricia, Hospital San Rafael, Madrid, España.
  • García-Cenador MB; Departamento de Cirugía, Universidad de Salamanca, Salamanca, España.
  • Mirón-Canelo JA; Departamento de Medicina Preventiva y Salud Pública, Universidad de Salamanca, Salamanca, España.
Actas Urol Esp ; 39(9): 558-63, 2015 Nov.
Article de En, Es | MEDLINE | ID: mdl-26072998
ABSTRACT

OBJECTIVES:

Since 1999, we have performed implantations of Remeex® adjustable prosthetics as rescue treatment for complex or persistent stress urinary incontinence (SUI) after failure of other surgical treatments (Burch colposuspension, Marshall-Marchetti-Krantz (MMK) or tension-free transvaginal tape [TVT] until 2003 and transobturator tape [TOT] since 2003). We present the results of our series, which include cases with severe complications. MATERIAL AND

METHOD:

Retrospective study of women diagnosed with complex or refractory SUI who underwent Remeex(®) implantation between October 1999 and December 2013. In 5 cases, we conducted cystocele correction in the same operation as the placement of the Remeex(®).

RESULTS:

Sixty women, with a mean age of 66.87 years (range 39-85), underwent operations. The procedure was successful in 68.33% of the cases. The failures consisted of the following 10% of the women had mixed urinary incontinence (UI) with multiple bladder diverticula; 8.33% had mild SUI; and 13.33% had urgency urination without UI. Thirty-five percent required adjustments. We recorded 3 cases with severe complications disabling severe UI in a patient who underwent multiple operations, massive pelvic hemorrhage in a patient undergoing standard antiplatelet therapy and infected vaginal calculi measuring 7cm on an extruding Remeex thread in a paraplegic patient 4 years after the implantation.

CONCLUSIONS:

Remeex(®) is an effective and safe procedure for achieving continence in cases of complex or refractory SUI, although it is not exempt from severe complications. Following rigorous protocols can help detect complications and treat them in a timely manner.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Complications postopératoires / Incontinence urinaire d'effort / Bandelettes sous-urétrales Type d'étude: Guideline / Observational_studies Limites: Aged / Aged80 / Female / Humans / Middle aged Langue: En / Es Journal: Actas Urol Esp Année: 2015 Type de document: Article

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Complications postopératoires / Incontinence urinaire d'effort / Bandelettes sous-urétrales Type d'étude: Guideline / Observational_studies Limites: Aged / Aged80 / Female / Humans / Middle aged Langue: En / Es Journal: Actas Urol Esp Année: 2015 Type de document: Article