Note: Directly measuring the direct digital synthesizer frequency chirp-rate for an atom interferometer.
Rev Sci Instrum
; 86(9): 096108, 2015 Sep.
Article
de En
| MEDLINE
| ID: mdl-26429495
During gravity measurements with Raman type atom interferometry, the frequency of the laser used to drive Raman transition is scanned by chirping the frequency of a direct digital synthesizer (DDS), and the local gravity is determined by precisely measuring the chip rate α of DDS. We present an effective method that can directly evaluate the frequency chirp rate stability of our DDS. By mixing a pair of synchronous linear sweeping signals, the chirp rate fluctuation is precisely measured with a frequency counter. The measurement result shows that the relative α instability can reach 5.7 × 10(-11) in 1 s, which is neglectable in a 10(-9) g level atom interferometry gravimeter.
Texte intégral:
1
Collection:
01-internacional
Base de données:
MEDLINE
Langue:
En
Journal:
Rev Sci Instrum
Année:
2015
Type de document:
Article
Pays de publication:
États-Unis d'Amérique