Your browser doesn't support javascript.
loading
Understanding the Scatter Radiation Distribution during C-arm CT Examination: A Body Phantom Study.
Norimasa, Toshiyo; Kakimi, Akihiko; Takao, Yoshinori; Sasaki, Shohei; Katayama, Yutaka; Himoto, Daisuke; Izuta, Shinichiro; Ichida, Takao.
Affiliation
  • Norimasa T; Department of Central Radiology, Osaka City University Hospital.
Nihon Hoshasen Gijutsu Gakkai Zasshi ; 72(11): 1144-1151, 2016.
Article de Ja | MEDLINE | ID: mdl-27867175
ABSTRACT
The purpose of this study was to understand the scatter radiation distribution during C-arm CT examination in the interventional radiography (IVR) room to show the escaped area and the radiation protective method. The C-arm rotates 200° in 5 s. The tube voltage was 90 kV, and the entrance dose to the detector was 0.36 µGy/frame during C-arm CT examination. The scattered doses were measured each 50 cm from the isocenter like a grid pattern. The heights of the measurement were 50, 100, and 150 cm from the floor. The maximum scattered doses were 38.23±0.60 µGy at 50 cm, 43.86±0.20 µGy at 100 cm, and 25.78±0.37 µGy at 150 cm. The scatter radiation distribution at 100 cm was the highest scattered dose. The operator should protect their reproductive gland, thyroid, and lens. The scattered dose was low behind the C-arm body and the bed, so they will be able to become the escaped area for staff.
Sujet(s)
Recherche sur Google
Collection: 01-internacional Base de données: MEDLINE Sujet principal: Tomodensitométrie / Fantômes en imagerie Langue: Ja Journal: Nihon Hoshasen Gijutsu Gakkai Zasshi Année: 2016 Type de document: Article
Recherche sur Google
Collection: 01-internacional Base de données: MEDLINE Sujet principal: Tomodensitométrie / Fantômes en imagerie Langue: Ja Journal: Nihon Hoshasen Gijutsu Gakkai Zasshi Année: 2016 Type de document: Article
...