Initial experience with a leadless pacemaker (Micra™) implantation in a low volume center in South East Asia.
Johar, Sofian; Luqman, Nazar.
Affiliation
  • Johar S; Consultant Cardiologist & Electrophysiologist, Gleneagles JPMC & Cardiac Centre RIPAS Hospital, Bander Seri Begawan BA1710, Brunei Darussalam.
  • Luqman N; Consultant Cardiologist, Cardiac Centre RIPAS Hospital, Bander Seri Begawan BA1710, Brunei Darussalam.
Future Cardiol ; 14(5): 389-395, 2018 Sep.
Article de En | MEDLINE | ID: mdl-30251546
Aim: The Micra™ Transcatheter Pacing System is a leadless pacemaker that has been introduced recently. We share our experience in a low volume center and the use of right ventricular angiography (RVA) during implantation. Materials & methods: Patients underwent Micra implantation and RVA was performed to predetermine the implant site.Results: Nine patients underwent Micra implantation. The most common indication was atrial fibrillation with bradycardia. The device was implanted at apical-septum in seven and mid-septum in two. The procedure time ranged from 30 to 100 min and fluoroscopic time 4-18 min. Pacing parameters remained stable after 1-month follow-up. Conclusion: The Micra implantation technique can be easily learnt. RVA was helpful in selecting an appropriate site for the Micra implant.
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Langue: En Journal: Future Cardiol Sujet du journal: CARDIOLOGIA Année: 2018 Type de document: Article Pays de publication: Royaume-Uni

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Langue: En Journal: Future Cardiol Sujet du journal: CARDIOLOGIA Année: 2018 Type de document: Article Pays de publication: Royaume-Uni