Your browser doesn't support javascript.
loading
Morse taper performance: A finite element analysis study.
Petris, Gustavo Passarelli; De Carli, João Paulo; Paranhos, Luiz Renato; Santos, Pâmela Letícia; Benetti, Paula; Walber, Marcio; Linden, Eduardo Sandini; Linden, Maria Salete Sandini.
Affiliation
  • Petris GP; Departamento de Odontologia, Universidade de Passo Fundo, Passo Fundo, RS, BR.
  • De Carli JP; Departamento de Odontologia, Universidade de Passo Fundo, Passo Fundo, RS, BR.
  • Paranhos LR; Departamento de Odontologia Preventiva e Social, Faculdade de Odontologia, Universidade Federal de Uberlandia (UFU), Uberlandia, MG, BR.
  • Santos PL; Departamento de Ciencias das Saude, Universidade de Araraquara (UNIARA), Araraquara, SP, BR.
  • Benetti P; Departamento de Odontologia, Universidade de Passo Fundo, Passo Fundo, RS, BR.
  • Walber M; Programa de Pos Graduacao em Engenharia Mecanica, Universidade de Passo Fundo, Passo Fundo, RS, BR.
  • Linden ES; Departamento de Odontologia, Universidade de Passo Fundo, Passo Fundo, RS, BR.
  • Linden MSS; Departamento de Odontologia, Universidade de Passo Fundo, Passo Fundo, RS, BR.
Clinics (Sao Paulo) ; 74: e852, 2019 03 25.
Article de En | MEDLINE | ID: mdl-30916212
ABSTRACT

OBJECTIVES:

To evaluate and compare the magnitude and distribution of stresses generated on implants, abutments and first molar metal-ceramic crowns using finite element analysis.

METHODS:

Preliminary three-dimensional models were created using the computer-aided design software SolidWorks. Stress and strain values were observed for two distinct virtual models model 1 - Morse taper and solid abutment; model 2 - Morse taper and abutment with screw. A load (250 N) was applied to a single point of the occlusal surface at 15° to the implant long axis. Von Mises stresses were recorded for both groups at four main points 1) abutment-retaining screws; 2) abutment neck; 3) cervical bone area; 4) implant neck. RESULTS AND

CONCLUSION:

Model 1 showed a higher stress value (1477.5 MPa) at the abutment-retaining screw area than the stresses found in model 2 (1091.1 MPa for the same area). The cervical bone strain values did not exceed 105 µm for either model.
Sujet(s)

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Implants dentaires / Conception de prothèse dentaire / Analyse des éléments finis / Analyse du stress dentaire / Conception d'implant dentaire et de pilier Limites: Humans Langue: En Journal: Clinics (Sao Paulo) Sujet du journal: MEDICINA Année: 2019 Type de document: Article Pays d'affiliation: Brésil

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Implants dentaires / Conception de prothèse dentaire / Analyse des éléments finis / Analyse du stress dentaire / Conception d'implant dentaire et de pilier Limites: Humans Langue: En Journal: Clinics (Sao Paulo) Sujet du journal: MEDICINA Année: 2019 Type de document: Article Pays d'affiliation: Brésil