Your browser doesn't support javascript.
loading
Clinical performance and sample freeze-thaw stability of the cobas®6800 SARS-CoV-2 assay for the detection of SARS-CoV-2 in oro-/nasopharyngeal swabs and lower respiratory specimens.
Stohr, Joep J J M; Wennekes, Meike; van der Ent, Martijn; Diederen, Bram M W; Kluytmans-van den Bergh, Marjolein F Q; Bergmans, Anneke M C; Kluytmans, Jan A J W; Pas, Suzan D.
Affiliation
  • Stohr JJJM; Department of Infection Control, Amphia Hospital, Breda, the Netherlands; Microvida Laboratory for Microbiology, Elisabeth-TweeSteden Hospital, Tilburg, the Netherlands; Microvida Laboratory for Microbiology, Amphia Hospital, Breda, the Netherlands. Electronic address: j.stohr@etz.nl.
  • Wennekes M; Microvida Laboratory for Microbiology, Bravis Hospital, Roosendaal, the Netherlands.
  • van der Ent M; Microvida Laboratory for Microbiology, Bravis Hospital, Roosendaal, the Netherlands.
  • Diederen BMW; Microvida Laboratory for Microbiology, Elisabeth-TweeSteden Hospital, Tilburg, the Netherlands; Microvida Laboratory for Microbiology, Amphia Hospital, Breda, the Netherlands; Microvida Laboratory for Microbiology, Bravis Hospital, Roosendaal, the Netherlands.
  • Kluytmans-van den Bergh MFQ; Department of Infection Control, Amphia Hospital, Breda, the Netherlands; Amphia Academy Infectious Disease Foundation, Amphia Hospital, Breda, the Netherlands; Julius Center for Health Sciences and Primary Care, University Medical Center Utrecht, Utrecht University, Utrecht, the Netherlands.
  • Bergmans AMC; Microvida Laboratory for Microbiology, Bravis Hospital, Roosendaal, the Netherlands.
  • Kluytmans JAJW; Department of Infection Control, Amphia Hospital, Breda, the Netherlands; Microvida Laboratory for Microbiology, Elisabeth-TweeSteden Hospital, Tilburg, the Netherlands; Microvida Laboratory for Microbiology, Amphia Hospital, Breda, the Netherlands; Amphia Academy Infectious Disease Foundation, Amph
  • Pas SD; Microvida Laboratory for Microbiology, Amphia Hospital, Breda, the Netherlands; Microvida Laboratory for Microbiology, Bravis Hospital, Roosendaal, the Netherlands.
J Clin Virol ; 133: 104686, 2020 12.
Article de En | MEDLINE | ID: mdl-33221622
ABSTRACT

INTRODUCTION:

Studies describing the performance characteristics of the cobas®6800 system for SARS-CoV-2 detection in deep respiratory specimens and freeze-thaw stability are limited. The current study compares the clinical performance of the automated SARS-CoV-2 assay on the cobas®6800 system to a lab-developed assay (LDA) and the cobas impact of freeze-thawing combined with lysis buffer.

METHODS:

Both retrospective and prospectively selected deep respiratory samples and oro- and nasopharyngeal samples in either E-swab® or GLY- were tested using the SARS-CoV-2 assay on the cobas®6800 System and compared to a lab developed assay. Additonally, SARS-CoV-2 RNA stability was assessed after one freeze-thaw cycle with or without lysis buffer.

RESULTS:

In total, 221 (58.3 %) oro- and nasopharyngeal swabs, 131 (34.6 %) deep respiratory specimens, and n = 25 (6.6 %) swabs of unknown origin were included to study clinical performance. Only 4 samples gave discrepant results, all being positive in the LDA and not the cobas®6800 system. For stability testing, 66 samples without and 110 with lysis buffer were included. No clinically significant difference was found in test results after one freeze-thaw cycle and addition of lysis buffer.

CONCLUSION:

Based on our findings, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results as the LDA in oro-/nasopharyngeal swabs and deep respiratory specimens. Moreover, the cobas®6800 SARS-CoV-2 RNA assay yielded similar results before and after a freeze-thaw cycle, with better preservation of low viral loads in lysis buffer.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Appareil respiratoire / Manipulation d'échantillons / Partie nasale du pharynx / Congélation / Détection de l'acide nucléique du virus de la COVID-19 / COVID-19 Type d'étude: Diagnostic_studies / Observational_studies Limites: Humans Langue: En Journal: J Clin Virol Sujet du journal: VIROLOGIA Année: 2020 Type de document: Article

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Appareil respiratoire / Manipulation d'échantillons / Partie nasale du pharynx / Congélation / Détection de l'acide nucléique du virus de la COVID-19 / COVID-19 Type d'étude: Diagnostic_studies / Observational_studies Limites: Humans Langue: En Journal: J Clin Virol Sujet du journal: VIROLOGIA Année: 2020 Type de document: Article