Your browser doesn't support javascript.
loading
Standardized artificially created stable pertrochanteric femur fractures present more homogenous results compared to osteotomies for orthopaedic implant testing.
Schader, J F; Zderic, I; Gehweiler, D; Dauwe, J; Mys, K; Danker, C; Acklin, Y P; Sommer, C; Gueorguiev, B; Stoffel, K.
Affiliation
  • Schader JF; AO Research Institute Davos, Clavadelerstrasse 8, 7270, Davos Platz, Switzerland. janaschader@gmail.com.
  • Zderic I; University of Basel, Basel, Switzerland. janaschader@gmail.com.
  • Gehweiler D; Department of Surgery, Cantonal Hospital Graubuenden, Chur, Switzerland. janaschader@gmail.com.
  • Dauwe J; AO Research Institute Davos, Clavadelerstrasse 8, 7270, Davos Platz, Switzerland.
  • Mys K; AO Research Institute Davos, Clavadelerstrasse 8, 7270, Davos Platz, Switzerland.
  • Danker C; AO Research Institute Davos, Clavadelerstrasse 8, 7270, Davos Platz, Switzerland.
  • Acklin YP; Department of Trauma Surgery, UZ Leuven, Leuven, Belgium.
  • Sommer C; AO Research Institute Davos, Clavadelerstrasse 8, 7270, Davos Platz, Switzerland.
  • Gueorguiev B; AO Research Institute Davos, Clavadelerstrasse 8, 7270, Davos Platz, Switzerland.
  • Stoffel K; University Hospital Basel, Basel, Switzerland.
BMC Musculoskelet Disord ; 22(1): 371, 2021 Apr 20.
Article de En | MEDLINE | ID: mdl-33879133
ABSTRACT

BACKGROUND:

With regard to biomechanical testing of orthopaedic implants, there is no consensus on whether artificial creation of standardized bone fractures or their simulation by means of osteotomies result in more realistic outcomes. Therefore, the aim of this study was to artificially create and analyze in an appropriate setting the biomechanical behavior of standardized stable pertrochanteric fractures versus their simulation via osteotomizing.

METHODS:

Eight pairs of fresh-frozen human cadaveric femora aged 72.7 ± 14.9 years (range 48-89 years) were assigned in paired fashion to two study groups. In Group 1, stable pertrochanteric fractures AO/OTA 31-A1 were artificially created via constant force application on the anterior cortex of the femur through a blunt guillotine blade. The same fracture type was simulated in Group 2 by means of osteotomies. All femora were implanted with a dynamic hip screw and biomechanically tested in 20° adduction under progressively increasing physiologic cyclic axial loading at 2 Hz, starting at 500 N and increasing at a rate of 0.1 N/cycle. Femoral head fragment movements with respect to the shaft were monitored by means of optical motion tracking.

RESULTS:

Cycles/failure load at 15° varus deformation, 10 mm leg shortening and 15° femoral head rotation around neck axis were 11324 ± 848/1632.4 ± 584.8 N, 11052 ± 1573/1605.2 ± 657.3 N and 11849 ± 1120/1684.9 ± 612.0 N in Group 1, and 10971 ± 2019/1597.1 ± 701.9 N, 10681 ± 1868/1568.1 ± 686.8 N and 10017 ± 4081/1501.7 ± 908.1 N in Group 2, respectively, with no significant differences between the two groups, p ≥ 0.233.

CONCLUSION:

From a biomechanical perspective, by resulting in more consistent outcomes under dynamic loading, standardized artificial stable pertrochanteric femur fracture creation may be more suitable for orthopaedic implant testing compared to osteotomizing the bone.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Orthopédie Limites: Aged / Aged80 / Humans / Middle aged Langue: En Journal: BMC Musculoskelet Disord Sujet du journal: FISIOLOGIA / ORTOPEDIA Année: 2021 Type de document: Article Pays d'affiliation: Suisse

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Orthopédie Limites: Aged / Aged80 / Humans / Middle aged Langue: En Journal: BMC Musculoskelet Disord Sujet du journal: FISIOLOGIA / ORTOPEDIA Année: 2021 Type de document: Article Pays d'affiliation: Suisse