Your browser doesn't support javascript.
loading
Profile of a new extended range-of-vision IOL: a laboratory study.
Tognetto, Daniele; Giglio, Rosa; De Giacinto, Chiara; Pastore, Marco R; Cirigliano, Gabriella; Piñero, David P; Turco, Gianluca.
Affiliation
  • Tognetto D; Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste, Italy.
  • Giglio R; Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste, Italy. giglio.rosam@gmail.com.
  • De Giacinto C; Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste, Italy.
  • Pastore MR; Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste, Italy.
  • Cirigliano G; Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste, Italy.
  • Piñero DP; Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, Spain.
  • Turco G; Department of Medicine, Surgery and Health Sciences, University of Trieste, Trieste, Italy.
Graefes Arch Clin Exp Ophthalmol ; 260(3): 913-916, 2022 Mar.
Article de En | MEDLINE | ID: mdl-34605952
ABSTRACT

PURPOSE:

To evaluate the surface profile of a new-generation extended range-of-vision intraocular lens (IOL) and to compare it with that obtained for a monofocal IOL based on the same platform.

METHODS:

Prospective, experimental, laboratory study comparing the surface profile of the DFT015 (AcrySof IQ Vivity; Alcon Laboratories, Inc.), a new-generation presbyopia-correcting IOL, with the profile of the SN60WF (AcrySof IQ; Alcon Laboratories, Inc.), an aspheric monofocal IOL based on the same platform. Raw profiles were obtained using contact profilometry. The best-fit form was then subtracted from each raw profile to highlight potential differences.

RESULTS:

No significant differences were appreciated in raw profiles. On the contrary, after form removal, the new extended range-of-vision IOL showed a peculiar profile characterized by the presence of two altitudinal symmetrical changes in the order of 1 µm, localized in the central portion of the optic.

CONCLUSIONS:

The new-generation extended range-of-vision IOL evaluated showed a smooth change of its surface compared to the same platform monofocal IOL. The altitudinal changes blended in the central design of the new presbyopia-correcting IOL, although micrometric, might play a crucial role in creating a continuous focal range while minimizing visual disturbances.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Phacoémulsification / Lentilles intraoculaires Type d'étude: Observational_studies Limites: Humans Langue: En Journal: Graefes Arch Clin Exp Ophthalmol Année: 2022 Type de document: Article Pays d'affiliation: Italie

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Phacoémulsification / Lentilles intraoculaires Type d'étude: Observational_studies Limites: Humans Langue: En Journal: Graefes Arch Clin Exp Ophthalmol Année: 2022 Type de document: Article Pays d'affiliation: Italie
...