High-flow nasal cannula therapy, factors affecting effective inspired oxygen fraction: an experimental adult bench model.
Duprez, Frédéric; de Terwangne, C; Bellemans, V; Poncin, W; Reychler, G; Sorgente, A; Cuvelier, G; Mashayekhi, S; Wittebole, X.
Affiliation
  • Duprez F; Department of Intensive Care, Centre Hospitalier Epicura, Hornu, Belgium. frederic.duprez@condorcet.be.
  • de Terwangne C; Laboratory of Respiratory Physiology, Condorcet School, Tournai, Belgium. frederic.duprez@condorcet.be.
  • Bellemans V; Department of Geriatric Medicine, Cliniques Universitaires Saint-Luc, Brussels, Belgium. Christophe_terwangne@hotmail.com.
  • Poncin W; , Brussels, Belgium. Christophe_terwangne@hotmail.com.
  • Reychler G; Department of Intensive Care, Centre Hospitalier Epicura, Hornu, Belgium.
  • Sorgente A; Department of Internal Medicine, Université Catholique de Louvain, Brussels, Belgium.
  • Cuvelier G; Department of Internal Medicine, Université Catholique de Louvain, Brussels, Belgium.
  • Mashayekhi S; Department of Cardiology, Centre Hospitalier Epicura, Hornu, Belgium.
  • Wittebole X; Laboratory of Respiratory Physiology, Condorcet School, Tournai, Belgium.
J Clin Monit Comput ; 36(5): 1441-1448, 2022 10.
Article de En | MEDLINE | ID: mdl-34877626
Oxygenation through High Flow Delivery Systems (HFO) is described as capable of delivering accurate FiO2. Meanwhile, peak inspiratory flow [Formula: see text] ) of patients with acute hypoxemic respiratory failure can reach up to 120 L/min, largely exceeding HFO flow. Currently, very few data on the reliability of HFO devices at these high [Formula: see text] are available. We sought to evaluate factors affecting oxygenation while using HFO systems at high [Formula: see text] in a bench study. Spontaneous breathing was generated with a mechanical test lung connected to a mechanical ventilator Servo-i®, set to volume control mode. Gas flow from a HFO device was delivered to the test lung. The influence on effective inspired oxygen fraction of three parameters (FiO2 0.6, 0.8, and 1, [Formula: see text] from 28 to 98.1 L/min, and HFO Gas Flows from 40 to 60 L/min) were analyzed and are reported. The present bench study demonstrates that during HFO treatment, measured FiO2 in the lung does not equal set FiO2 on the device. The substance of this variation (ΔFiO2) is tightly correlated to [Formula: see text] (Pearson's coefficient of 0.94, p-value < 0.001). Additionally, set FiO2 and Flow at HFO device appear to significatively affect ΔFiO2 as well (p-values < 0.001, adjusted to [Formula: see text] ). The result of multivariate linear regression indicates predictors ([Formula: see text] , Flow and set FiO2) to explain 92% of the variance of delta FiO2 through K-Fold Cross Validation. Moreover, adjunction of a dead space in the breathing circuit significantly decreased ΔFiO2 (p < 0.01). The present bench study did expose a weakness of HFO devices in reliability of delivering accurate FIO2 at high [Formula: see text] as well as, to a lesser extent, at [Formula: see text] below equivalent set HFO Flows. Moreover, set HFO flow and set FIO2 did influence the variability of effective inspired oxygen fraction. The adjunction of a dead space in the experimental set-up significantly amended this variability and should thus be further studied in order to improve success rate of HFO therapy.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Insuffisance respiratoire / Canule Type d'étude: Clinical_trials / Prognostic_studies Limites: Adult / Humans Langue: En Journal: J Clin Monit Comput Sujet du journal: INFORMATICA MEDICA / MEDICINA Année: 2022 Type de document: Article Pays d'affiliation: Belgique Pays de publication: Pays-Bas

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Insuffisance respiratoire / Canule Type d'étude: Clinical_trials / Prognostic_studies Limites: Adult / Humans Langue: En Journal: J Clin Monit Comput Sujet du journal: INFORMATICA MEDICA / MEDICINA Année: 2022 Type de document: Article Pays d'affiliation: Belgique Pays de publication: Pays-Bas