Your browser doesn't support javascript.
loading
Risk factors for poor performance in finger cuff non-invasive monitoring of arterial pressure: A prospective multicenter study.
Lakhal, Karim; Dauvergne, Jérôme E; Messet-Charriere, Hélène; Nay, Mai-Anh; Kamel, Toufik; Muller, Grégoire; Robert-Edan, Vincent; Rozec, Bertrand; Ehrmann, Stephan; Jacquier, Sophie; Boulain, Thierry.
Affiliation
  • Lakhal K; Nantes Université, CHU Nantes, Department of Anesthesiology and Critical Care at Laënnec Hospital, University Hospital of Nantes, F-44000 Nantes, France. Electronic address: lakhal_karim@yahoo.fr.
  • Dauvergne JE; Nantes Université, CHU Nantes, Department of Anesthesiology and Critical Care at Laënnec Hospital, University Hospital of Nantes, F-44000 Nantes, France.
  • Messet-Charriere H; CHRU Tours, Médecine Intensive Réanimation, CIC INSERM 1415, Tours, France.
  • Nay MA; Service de médecine intensive-réanimation, Centre Hospitalier Universitaire d'Orléans, 14 avenue de l'hôpital, 45100 Orléans, France.
  • Kamel T; Service de médecine intensive-réanimation, Centre Hospitalier Universitaire d'Orléans, 14 avenue de l'hôpital, 45100 Orléans, France.
  • Muller G; Service de médecine intensive-réanimation, Centre Hospitalier Universitaire d'Orléans, 14 avenue de l'hôpital, 45100 Orléans, France; CRICS-TriggerSep F-CRIN research network, Tours, France; Université de Tours, EA4245, Transplantation, Immunologie, Inflammation, Tours, France.
  • Robert-Edan V; Nantes Université, CHU Nantes, Department of Anesthesiology and Critical Care at Laënnec Hospital, University Hospital of Nantes, F-44000 Nantes, France.
  • Rozec B; Nantes Université, CHU Nantes, Department of Anesthesiology and Critical Care at Laënnec Hospital, University Hospital of Nantes, F-44000 Nantes, France; Institut du Thorax, Institut National de la Santé et de la Recherche Médicale (INSERM), Centre National de la Recherche Scientifique (CNRS), Unive
  • Ehrmann S; CHRU Tours, Médecine Intensive Réanimation, CIC INSERM 1415, Tours, France; CRICS-TriggerSep F-CRIN research network, Tours, France; INSERM, Centre d'étude des pathologies respiratoires, U1100 Tours, France; Université de Tours, Tours, France.
  • Jacquier S; CHRU Tours, Médecine Intensive Réanimation, CIC INSERM 1415, Tours, France.
  • Boulain T; Service de médecine intensive-réanimation, Centre Hospitalier Universitaire d'Orléans, 14 avenue de l'hôpital, 45100 Orléans, France.
Anaesth Crit Care Pain Med ; 43(2): 101333, 2024 Apr.
Article de En | MEDLINE | ID: mdl-38048986
ABSTRACT

BACKGROUND:

Compared to the invasive technique, non-invasive monitoring of arterial pressure favors easier and faster implementation while potentially sacrificing some reliability. This may be particularly true for the Clearsight™ system (Edwards Lifesciences), which enables continuous monitoring. We evaluated the risk factors for its poor performance.

METHODS:

Patients with an arterial catheter and stable mean arterial pressure (MAP) over a 5-min period were included. Six pairs of invasive and Clearsight measurements of MAP were collected and the bias between the two techniques was calculated. Poor performance of the Clearsight™ system was defined as either a failure to measure and display MAP or displaying an erroneous MAP (individual bias > 5 mmHg). Fingertip perfusion was assessed using the plethysmographic perfusion index (PI) and the capillary refill time (CRT).

RESULTS:

Among 152 ICU patients (MAP of 81 ± 14 mmHg, norepinephrine in 78 [51%]), 78 (51%) experienced a poor performance of the Clearsight™ system failure to display MAP in 19 (13%) patients, and erroneous value displayed in 59 (44%). In multivariate analysis, PI ≤ 0.85% (adjusted odds ratio [aOR] = 2.94 [95% confidence interval (95%CI)1.34;6.45]), CRT > 4 s (aOR = 5.28 [95%CI 1.39;20.05]), and the presence of hand edema (aOR = 2.06 [95%CI 1.01;4.21]) were associated with a higher likelihood of poor performance. Cardiac arrhythmia (aOR = 1.39 [95%CI 0.64;3.02]) and other tested variables were not associated with poor performance.

CONCLUSIONS:

Half of the included patients exhibited poor Clearsight™ system performance. Our results caution against using finger cuff arterial pressure monitoring in patients with low PI (≤0.85%), protracted CRT (>4 s), or hand edema. REGISTRATION ClinicalTrials.gov, NCT04269382, Dr. G. Muller, February 13, 2020. https//classic. CLINICALTRIALS gov/ct2/show/NCT04269382.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Oedème / Pression artérielle Limites: Humans Langue: En Journal: Anaesth Crit Care Pain Med Année: 2024 Type de document: Article

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Oedème / Pression artérielle Limites: Humans Langue: En Journal: Anaesth Crit Care Pain Med Année: 2024 Type de document: Article
...