[Accuracy of Injection Dose of Amyloid PET Agent Using Radiopharmaceutical Activity Suppliers].
Maeda, Yukito; Matsumoto, Keiichi; Ikari, Yasuhiko; Akamatsu, Go; Shimizu, Keiji; Tsuda, Keisuke.
Affiliation
  • Maeda Y; Department of Medical Technology, Kagawa University Hospital.
  • Matsumoto K; Department of Radiological Technology, Faculty of Medical Science, Kyoto College of Medical Science.
  • Ikari Y; Department of Molecular Imaging Research, Center of Clinical Research and Innovation, Kobe City Medical Center General Hospital.
  • Akamatsu G; Department of Molecular Imaging Research, Center of Clinical Research and Innovation, Kobe City Medical Center General Hospital.
  • Shimizu K; Department of Medical Physics and Engineering Course of Health Science, Osaka University Graduate School of Medicine.
  • Tsuda K; Department of Molecular Imaging Research, Center of Clinical Research and Innovation, Kobe City Medical Center General Hospital.
Nihon Hoshasen Gijutsu Gakkai Zasshi ; 80(2): 155-165, 2024 Feb 20.
Article de Ja | MEDLINE | ID: mdl-38072451
ABSTRACT

PURPOSE:

This study aimed to identify disposable items with low amyloid positron emission tomography (PET) agent radioactivity adsorption for accurate injections using a radiopharmaceutical activity supplier.

METHODS:

First, we investigated disposable items currently used for amyloid PET agent injection. Next, we measured the residual radioactivity rates of amyloid PET agents on three-way stopcocks, extension tubes, butterfly needles, and indwelling needles to identify disposable items with low radioactivity adsorption. Finally, we evaluated the accuracy of amyloid PET agent injection using the selected disposable items and a radiopharmaceutical activity supplier.

RESULTS:

The polybutadiene extension tube exhibited a significantly lower residual activity rate than that of the polyvinyl chloride extension tube. Similarly, the indwelling needles showed significantly lower residual activity rate than that of butterfly needles. The dose indicated by a radiopharmaceutical activity supplier was 184.1 MBq, while the dose calibrator measured the radioactivity which flowed into the vial as 170.2 MBq, resulting in an administration accuracy of 8.2%.

CONCLUSION:

To ensure accurate amyloid PET agent injections, we recommend using polybutadiene extension tubes and indwelling needles due to their lower radioactivity adsorption.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Radiopharmaceutiques / Élastomères / Tomographie par émission de positons Langue: Ja Journal: Nihon Hoshasen Gijutsu Gakkai Zasshi Année: 2024 Type de document: Article

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Radiopharmaceutiques / Élastomères / Tomographie par émission de positons Langue: Ja Journal: Nihon Hoshasen Gijutsu Gakkai Zasshi Année: 2024 Type de document: Article