Your browser doesn't support javascript.
loading
Determining the Aethalometer multiple scattering enhancement factor C from the filter loading parameter.
Ferrero, L; Losi, N; Rigler, M; Gregoric, A; Colombi, C; D'Angelo, L; Cuccia, E; Cefalì, A M; Gini, I; Doldi, A; Cerri, S; Maroni, P; Cipriano, D; Markuszewski, P; Bolzacchini, E.
Affiliation
  • Ferrero L; GEMMA and POLARIS Centre, Università degli Studi di Milano-Bicocca, 20126 Milano, Italy. Electronic address: luca.ferrero@unimib.it.
  • Losi N; GEMMA and POLARIS Centre, Università degli Studi di Milano-Bicocca, 20126 Milano, Italy.
  • Rigler M; Aerosol d.o.o., Kamniska 39A, SI-1000 Ljubljana, Slovenia.
  • Gregoric A; Aerosol d.o.o., Kamniska 39A, SI-1000 Ljubljana, Slovenia; Center for Atmospheric Research, University of Nova Gorica, SI-5000 Nova Gorica, Slovenia.
  • Colombi C; Regional Agency for Environmental Protection of Lombardy (ARPA Lombardia), Air Quality Department, Milan, Italy.
  • D'Angelo L; Regional Agency for Environmental Protection of Lombardy (ARPA Lombardia), Air Quality Department, Milan, Italy; Institute for Atmospheric and Environmental Sciences, Goethe-University Frankfurt, Frankfurt am Main 60438, Germany.
  • Cuccia E; Regional Agency for Environmental Protection of Lombardy (ARPA Lombardia), Air Quality Department, Milan, Italy.
  • Cefalì AM; GEMMA and POLARIS Centre, Università degli Studi di Milano-Bicocca, 20126 Milano, Italy; RSE - Ricerca sul Sistema Energetico S.p.A., via Rubattino 54, 20134 Milano, Italy.
  • Gini I; GEMMA and POLARIS Centre, Università degli Studi di Milano-Bicocca, 20126 Milano, Italy.
  • Doldi A; GEMMA and POLARIS Centre, Università degli Studi di Milano-Bicocca, 20126 Milano, Italy.
  • Cerri S; GEMMA and POLARIS Centre, Università degli Studi di Milano-Bicocca, 20126 Milano, Italy.
  • Maroni P; GEMMA and POLARIS Centre, Università degli Studi di Milano-Bicocca, 20126 Milano, Italy.
  • Cipriano D; RSE - Ricerca sul Sistema Energetico S.p.A., via Rubattino 54, 20134 Milano, Italy.
  • Markuszewski P; Institute of Oceanology, Polish Academy of Sciences, 81-712 Sopot, Poland; Bolin Centre for Climate Research and Department of Environmental Science, Stockholm University, 10691 Stockholm, Sweden.
  • Bolzacchini E; GEMMA and POLARIS Centre, Università degli Studi di Milano-Bicocca, 20126 Milano, Italy.
Sci Total Environ ; 917: 170221, 2024 Mar 20.
Article de En | MEDLINE | ID: mdl-38280585
ABSTRACT
Light-absorbing aerosols heat the atmosphere; an accurate quantification of their absorption coefficient is mandatory. However, standard reference instruments (CAPS, MAAP, PAX, PTAAM) are not always available at each measuring site around the world. By integrating all previous published studies concerning the Aethalometers, the AE33 filter loading parameter, provided by the dual-spot algorithm, were used to determine the multiple scattering enhancement factor from the Aethalometer itself (hereinafter CAE) on an yearly and a monthly basis. The method was developed in Milan, where Aethalometer measurements were compared with MAAP data; the comparison showed a good agreement in terms of equivalent black carbon (R2 = 0.93; slope = 1.02 and a negligible intercept = 0.12 µg m-3) leading to a yearly experimental multiple scattering enhancement factor of 2.51 ± 0.04 (hereinafter CMAAP). On a yearly time base the CAE values obtained using the new approach was 2.52 ± 0.01, corresponding to the experimental one (CMAAP). Considering the seasonal behavior, higher experimental CMAAP and computed CAE values were found in summer (2.83 ± 0.12) whereas, the lower ones in winter/early-spring (2.37 ± 0.03), in agreement with the single scattering albedo behavior in the Po Valley. Overall, the agreement between the experimental CMAAP and CAE showed a root mean squared error (RMSE) of just 0.038 on the CMAAP prediction, characterized by a slope close to 1 (1.001 ± 0.178), a negligible intercept (-0.002 ± 0.455) and a high degree of correlation (R2 = 0.955). From an environmental point of view, the application of a dynamic (space/time) determination of CAE increases the accuracy of the aerosol heating rate (compared to applying a fixed C value) up to 16 % solely in Milan, and to 114 % when applied in the Arctic at 80°N.