Your browser doesn't support javascript.
loading
Primary stability in locking plate fixation for proximal humeral fractures may be increased by using an additional nail osteosynthesis in combination with locking plate osteosynthesis-A biomechanical comparison.
Schöbel, Tobias; Gemkow, Maximilian; Wendler, Toni; Schleifenbaum, Stefan; Löffler, Sabine; Theopold, Jan; Hepp, Pierre.
Affiliation
  • Schöbel T; Department of Orthopedic, Trauma, and Plastic Surgery, University of Leipzig, Liebigstraße 20, 04103 Leipzig, Germany. Electronic address: tobias.schoebel@medizin.uni-leipzig.de.
  • Gemkow M; Department of Orthopedic, Trauma, and Plastic Surgery, University of Leipzig, Liebigstraße 20, 04103 Leipzig, Germany. Electronic address: maximilian.gemkow@medizin.uni-leipzig.de.
  • Wendler T; Department of Orthopedic, Trauma, and Plastic Surgery, University of Leipzig, Liebigstraße 20, 04103 Leipzig, Germany; ZESBO - Center for Research on Musculoskeletal Systems, Semmelweisstraße 14, 04103 Leipzig, Germany. Electronic address: toni.wendler@medizin.uni-leipzig.de.
  • Schleifenbaum S; Department of Orthopedic, Trauma, and Plastic Surgery, University of Leipzig, Liebigstraße 20, 04103 Leipzig, Germany; ZESBO - Center for Research on Musculoskeletal Systems, Semmelweisstraße 14, 04103 Leipzig, Germany; Fraunhofer Institute for Machine Tools and Forming Technology, Nöthnitzer Straße
  • Löffler S; Institute of Anatomy, University of Leipzig, Liebigstraße 13, 04103 Leipzig, Germany. Electronic address: sabine.loeffler@medizin.uni-leipzig.de.
  • Theopold J; Department of Orthopedic, Trauma, and Plastic Surgery, University of Leipzig, Liebigstraße 20, 04103 Leipzig, Germany. Electronic address: jan.theopold@medizin.uni-leipzig.de.
  • Hepp P; Department of Orthopedic, Trauma, and Plastic Surgery, University of Leipzig, Liebigstraße 20, 04103 Leipzig, Germany. Electronic address: pierre.hepp@medizin.uni-leipzig.de.
Clin Biomech (Bristol, Avon) ; 114: 106235, 2024 Apr.
Article de En | MEDLINE | ID: mdl-38552372
ABSTRACT

BACKGROUND:

Stability is regarded as an important factor for proper healing and avoiding secondary dislocation in osteoporotic fractures of the proximal humerus. Several surgical techniques have been proposed for treatment, including intramedullary nail osteosynthesis and locking plate fixation. This study introduces a novel approach that combines both techniques and compares its primary stability with conventional methods.

METHODS:

The study involved 25 osteoporotic humeri with two-part fracture models, which were randomly assigned to locking-plate fixation, intramedullary nailing, or a combination of both techniques. The specimens were subjected to sinusoidal loading at 250 N in 20° abduction for 5000 cycles and then to quasi-static loading until failure. Fracture movement, failure mode, and failure load were measured and compared among the groups.

FINDINGS:

The groups fixated with intramedullary nailing and the groups fixated with intramedullary nailing and locking plate fixation showed significantly lower fracture motion than the group using locking plate fixation only (p < 0.005) and significantly higher load to failure (p = 0.007 and p = 0.0062, respectively). There was no significant difference between the group using intramedullary nailing and the group using locking-plate fixation and intramedullary nailing in fracture movement or load-to-failure (p > 0.005).

INTERPRETATION:

The results indicate that locking plate fixation provides less primary stability than intramedullary nailing or the combined of both techniques. This combined approach may offer advantages as a treatment for complex proximal humeral fractures in osteoporotic bone, and specific implants should be developed to ensure optimal treatment.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Fractures de l&apos;épaule / Fractures ostéoporotiques / Ostéosynthese intramedullaire / Fractures de l&apos;humérus Limites: Humans Langue: En Journal: Clin Biomech (Bristol, Avon) Sujet du journal: ENGENHARIA BIOMEDICA / FISIOLOGIA Année: 2024 Type de document: Article

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Fractures de l&apos;épaule / Fractures ostéoporotiques / Ostéosynthese intramedullaire / Fractures de l&apos;humérus Limites: Humans Langue: En Journal: Clin Biomech (Bristol, Avon) Sujet du journal: ENGENHARIA BIOMEDICA / FISIOLOGIA Année: 2024 Type de document: Article