Radiation dose estimation with multiple artificial neural networks in dicentric chromosome assay.
Jang, Seungsoo; Lee, Janghee; Kim, Song-Hyun; Han, Sangsoo; Shin, Sung-Gyun; Lee, Sunghee; Kang, Inhyuk; Jo, Wol Soon; Jeong, Sookyung; Oh, Su Jung; Lee, Chang Geun.
Affiliation
  • Jang S; Department of Advanced Nuclear Engineering, POSTECH, Pohang, Korea.
  • Lee J; Department of Advanced Nuclear Engineering, POSTECH, Pohang, Korea.
  • Kim SH; SierraBASE Co. Ltd, Sejong, Korea.
  • Han S; SierraBASE Co. Ltd, Sejong, Korea.
  • Shin SG; SierraBASE Co. Ltd, Sejong, Korea.
  • Lee S; SierraBASE Co. Ltd, Sejong, Korea.
  • Kang I; SierraBASE Co. Ltd, Sejong, Korea.
  • Jo WS; Research Center, Dongnam Institute of Radiological and Medical Science, Busan, Korea.
  • Jeong S; Research Center, Dongnam Institute of Radiological and Medical Science, Busan, Korea.
  • Oh SJ; Research Center, Dongnam Institute of Radiological and Medical Science, Busan, Korea.
  • Lee CG; Research Center, Dongnam Institute of Radiological and Medical Science, Busan, Korea.
Int J Radiat Biol ; 100(6): 865-874, 2024.
Article de En | MEDLINE | ID: mdl-38687685
ABSTRACT

PURPOSE:

The dicentric chromosome assay (DCA), often referred to as the 'gold standard' in radiation dose estimation, exhibits significant challenges as a consequence of its labor-intensive nature and dependency on expert knowledge. Existing automated technologies face limitations in accurately identifying dicentric chromosomes (DCs), resulting in decreased precision for radiation dose estimation. Furthermore, in the process of identifying DCs through automatic or semi-automatic methods, the resulting distribution could demonstrate under-dispersion or over-dispersion, which results in significant deviations from the Poisson distribution. In response to these issues, we developed an algorithm that employs deep learning to automatically identify chromosomes and perform fully automatic and accurate estimation of diverse radiation doses, adhering to a Poisson distribution. MATERIALS AND

METHODS:

The dataset utilized for the dose estimation algorithm was generated from 30 healthy donors, with samples created across seven doses, ranging from 0 to 4 Gy. The procedure encompasses several

steps:

extracting images for dose estimation, counting chromosomes, and detecting DC and fragments. To accomplish these tasks, we utilize a diverse array of artificial neural networks (ANNs). The identification of DCs was accomplished using a detection mechanism that integrates both deep learning-based object detection and classification methods. Based on these detection results, dose-response curves were constructed. A dose estimation was carried out by combining a regression-based ANN with the Monte-Carlo method.

RESULTS:

In the process of extracting images for dose analysis and identifying DCs, an under-dispersion tendency was observed. To rectify the discrepancy, classification ANN was employed to identify the results of DC detection. This approach led to satisfaction of Poisson distribution criteria by 32 out of the initial pool of 35 data points. In the subsequent stage, dose-response curves were constructed using data from 25 donors. Data provided by the remaining five donors served in performing dose estimations, which were subsequently calibrated by incorporating a regression-based ANN. Of the 23 points, 22 fell within their respective confidence intervals at p < .05 (95%), except for those associated with doses at levels below 0.5 Gy, where accurate calculation was obstructed by numerical issues. The accuracy of dose estimation has been improved for all radiation levels, with the exception of 1 Gy.

CONCLUSIONS:

This study successfully demonstrates a high-precision dose estimation method across a general range up to 4 Gy through fully automated detection of DCs, adhering strictly to Poisson distribution. Incorporating multiple ANNs confirms the ability to perform fully automated radiation dose estimation. This approach is particularly advantageous in scenarios such as large-scale radiological incidents, improving operational efficiency and speeding up procedures while maintaining consistency in assessments. Moreover, it reduces potential human error and enhances the reliability of results.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Dose de rayonnement / Aberrations des chromosomes / 29935 Limites: Humans Langue: En Journal: Int J Radiat Biol Sujet du journal: RADIOLOGIA Année: 2024 Type de document: Article Pays de publication: Royaume-Uni

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Dose de rayonnement / Aberrations des chromosomes / 29935 Limites: Humans Langue: En Journal: Int J Radiat Biol Sujet du journal: RADIOLOGIA Année: 2024 Type de document: Article Pays de publication: Royaume-Uni