Improving diagnostic confidence in low-dose dual-energy CTE with low energy level and deep learning reconstruction.
Lin, Xu; Gao, Yankun; Zhu, Chao; Song, Jian; Liu, Ling; Li, Jianying; Wu, Xingwang.
Affiliation
  • Lin X; Department of Radiology, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui 230022, China.
  • Gao Y; Department of Radiology, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui 230022, China.
  • Zhu C; Department of Radiology, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui 230022, China.
  • Song J; Department of Radiology, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui 230022, China.
  • Liu L; CT Research Center, GE Healthcare China, Shanghai 210000, China.
  • Li J; CT Research Center, GE Healthcare China, Shanghai 210000, China.
  • Wu X; Department of Radiology, The First Affiliated Hospital of Anhui Medical University, Hefei, Anhui 230022, China. Electronic address: duobi2004@126.com.
Eur J Radiol ; 178: 111607, 2024 Sep.
Article de En | MEDLINE | ID: mdl-39033690
ABSTRACT

OBJECTIVE:

To demonstrate the value of using 50 keV virtual monochromatic images with deep learning image reconstruction (DLIR) in low-dose dual-energy CT enterography (CTE).

METHODS:

In this prospective study, 114 participants (62 % M; 41.9 ± 16 years) underwent dual-energy CTE. The early-enteric phase was performed using standard-dose (noise index (NI) 8) and images were reconstructed at 70 keV and 50 keV with 40 % strength ASIR-V (ASIR-V40%). The late-enteric phase used low-dose (NI 12) and images were reconstructed at 50 keV with ASIR-V40%, and DLIR at medium (DLIR-M) and high strength (DLIR-H). Image standard deviation (SD), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), edge-rise-slope (ERS) were computed. The quantitative comb sign score was calculated for the 27 patients with Crohn's disease. The subjective noise, image contrast, display of rectus artery were scored using a 5-point scale by two radiologists blindly.

RESULTS:

Effective dose was reduced by 50 % (P < 0.001) in the late-enteric phase to 3.26 mSv. The lower-dose 50 keV-DLIR-H images (SD17.7 ± 0.5HU) had similar image noise (P = 0.97) as the standard-dose 70 keV-ASIR-V40% images (SD17.7 ± 0.73HU), but with higher (P < 0.001) SNR, CNR, ERS and quantitative comb sign score (5.7 ± 0.17, 1.8 ± 0.12, 156.04 ± 5.21 and 5.05 ± 0.73, respectively). Furthermore, the lower-dose 50 keV-DLIR-H images obtained the highest score in the rectus artery visibility (4.27 ± 0.6).

CONCLUSIONS:

The 50 keV images in dual-energy CTE with DLIR provides high-quality images, with a 50 % reduction in radiation dose. Images with high contrast and density resolutions significantly enhance the diagnostic confidence of Crohn's disease and are essential for the clinical development of individualized treatment plans.
Sujet(s)
Mots clés

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Dose de rayonnement / Interprétation d&apos;images radiographiques assistée par ordinateur / Tomodensitométrie / Radiographie digitale par projection en double énergie / Apprentissage profond Limites: Adult / Aged / Female / Humans / Male / Middle aged Langue: En Journal: Eur J Radiol Année: 2024 Type de document: Article Pays d'affiliation: Chine

Texte intégral: 1 Collection: 01-internacional Base de données: MEDLINE Sujet principal: Dose de rayonnement / Interprétation d&apos;images radiographiques assistée par ordinateur / Tomodensitométrie / Radiographie digitale par projection en double énergie / Apprentissage profond Limites: Adult / Aged / Female / Humans / Male / Middle aged Langue: En Journal: Eur J Radiol Année: 2024 Type de document: Article Pays d'affiliation: Chine