Determination of capsaicin and zucapsaicin in human serum by high-performance liquid chromatography with fluorescence detection.
Lu, J; Cwik, M.
Affiliation
  • Lu J; Clinical Research Laboratory, College of Pharmacy, University of Illinois at Chicago, 60612, USA.
J Chromatogr B Biomed Sci Appl ; 701(1): 135-9, 1997 Nov 07.
Article de En | MEDLINE | ID: mdl-9389349
ABSTRACT
A reversed-phase high-performance liquid chromatographic (HPLC) assay was developed to analyze capsaicin and zucapsaicin (civamide) in human serum at concentrations from 1 to 100 ng/ml. Human serum specimens were extracted twice with hexane-methyl tert.-butyl ether (11). The chromatographic separation was carried out on a C18 column at 40 degrees C using a mobile phase consisting of 40% acetonitrile in water with 5% tetrahydrofuran and 1% acetic acid. The concentration of the eluting compounds was monitored by a fluorescence detector with excitation at 270 nm and an emission cutoff of 300 nm. No interferences were observed from the extract of blank serum. The standard curves were linear in the detection range. The relative standard deviation of the assay was better than 8.4%. The limit of detection was 0.5 ng/ml.
Sujet(s)
Recherche sur Google
Collection: 01-internacional Base de données: MEDLINE Sujet principal: Capsaïcine / Analgésiques non narcotiques Type d'étude: Diagnostic_studies Limites: Humans Langue: En Journal: J Chromatogr B Biomed Sci Appl Sujet du journal: QUIMICA CLINICA Année: 1997 Type de document: Article Pays d'affiliation: États-Unis d'Amérique
Recherche sur Google
Collection: 01-internacional Base de données: MEDLINE Sujet principal: Capsaïcine / Analgésiques non narcotiques Type d'étude: Diagnostic_studies Limites: Humans Langue: En Journal: J Chromatogr B Biomed Sci Appl Sujet du journal: QUIMICA CLINICA Année: 1997 Type de document: Article Pays d'affiliation: États-Unis d'Amérique
...