Response of a tissue equivalent proportional counter to neutrons.
Badhwar, G D; Robbins, D E; Gibbons, F; Braby, L A.
Affiliation
  • Badhwar GD; NASA Johnson Space Center, Houston, TX 77058-3696, USA.
Radiat Meas ; 35(6): 551-6, 2002 Dec.
Article in En | MEDLINE | ID: mdl-12455500
The absorbed dose as a function of lineal energy was measured at the CERN-EC Reference-field Facility (CERF) using a 512-channel tissue equivalent proportional counter (TEPC), and neutron dose equivalent response evaluated. Although there are some differences, the measured dose equivalent is in agreement with that measured by the 16-channel HANDI tissue equivalent counter. Comparison of TEPC measurements with those made by a silicon solid-state detector for low linear energy transfer particles produced by the same beam, is presented. The measurements show that about 4% of dose equivalent is delivered by particles heavier than protons generated in the conducting tissue equivalent plastic.
Subject(s)
Key words
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Radiation Monitoring / Spacecraft / Linear Energy Transfer / Neutrons Language: En Journal: Radiat Meas Journal subject: RADIOLOGIA Year: 2002 Document type: Article Affiliation country: Country of publication: