Your browser doesn't support javascript.
loading
Modular prosthetic reconstruction of major bone defects of the distal end of the humerus.
Funovics, Philipp T; Schuh, Reinhard; Adams, Samuel B; Sabeti-Aschraf, Manu; Dominkus, Martin; Kotz, Rainer I.
Affiliation
  • Funovics PT; Department of Orthopaedics, Medical University of Vienna, Vienna General Hospital, Waehringer Guertel 18-20, 1090 Vienna, Austria. philipp.funovics@meduniwien.ac.at
J Bone Joint Surg Am ; 93(11): 1064-74, 2011 Jun 01.
Article in En | MEDLINE | ID: mdl-21655900
ABSTRACT

BACKGROUND:

Bone defects of the distal end of the humerus require complex reconstructions, for which standard prostheses may be insufficient. We investigated the outcomes of distal humeral reconstruction with use of a modular prosthesis.

METHODS:

Fifty-three elbows in fifty-two patients underwent reconstruction with a modular prosthesis (twelve total humeral replacements and forty-one distal humeral replacements) after tumor resection (thirty-eight elbows) or because of massive joint degeneration (fifteen elbows). In the tumor group, twenty-three patients (twenty-four elbows) had metastatic disease and fourteen had a primary tumor. Degenerative defects of the distal end of the humerus were caused by pseudarthrosis (six elbows), prosthetic failure (five), trauma (two), osteomyelitis (one), and supracondylar fracture (one). The mean duration of follow-up for all patients was twenty-eight months (median, thirteen months; range, one to 219 months).

RESULTS:

The mean Inglis-Pellicci score in the tumor group was 84 points, and the mean Musculoskeletal Tumor Society score was 78%. Patients with total humeral reconstruction had worse scores than those with distal humeral reconstruction. Twenty-four patients died of disease at a mean of thirteen months after surgery. Local tumor control was achieved in all patients. In the revision group, the mean Inglis-Pellicci score was 76 points. The Inglis-Pellicci score was significantly better for patients in the tumor group. Eight patients (15%) had a deep periprosthetic infection, requiring amputation in one patient (2%) and prosthetic removal in two patients (4%). Four patients (8%) had the implants revised for aseptic loosening.

CONCLUSIONS:

Modular prostheses of the distal end of the humerus provide a stable reconstruction of the elbow with satisfactory function and disease control in patients with a tumor, but careful patient selection is required when the prostheses are used for revision surgery in patients without a tumor.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Arthroplasty, Replacement / Prosthesis Implantation / Plastic Surgery Procedures / Humerus / Joint Prosthesis Limits: Adolescent / Adult / Aged / Aged80 / Child / Female / Humans / Male / Middle aged Language: En Journal: J Bone Joint Surg Am Year: 2011 Document type: Article Affiliation country:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Arthroplasty, Replacement / Prosthesis Implantation / Plastic Surgery Procedures / Humerus / Joint Prosthesis Limits: Adolescent / Adult / Aged / Aged80 / Child / Female / Humans / Male / Middle aged Language: En Journal: J Bone Joint Surg Am Year: 2011 Document type: Article Affiliation country: