Your browser doesn't support javascript.
loading
Comparison of trueness and repeatability of facial prosthesis design using a 3D morphable model approach, traditional computer-aided design methods, and conventional manual sculpting techniques.
Jablonski, Rachael Y; Malhotra, Taran; Shaw, Daniel; Coward, Trevor J; Shuweihdi, Farag; Bojke, Chris; Pavitt, Sue H; Nattress, Brian R; Keeling, Andrew J.
Affiliation
  • Jablonski RY; Specialty Registrar in Restorative Dentistry and NIHR Doctoral Fellow, Department of Restorative Dentistry, School of Dentistry, University of Leeds, Leeds, England, UK. Electronic address: rachaeljablonski@gmail.com.
  • Malhotra T; Lead Specialist Maxillofacial Prosthetist, Maxillofacial Prosthetics Laboratory, Liverpool University Hospitals NHS Foundation Trust, Aintree University Hospital, Liverpool, England, UK.
  • Shaw D; Maxillofacial Laboratory Manager, Maxillofacial Department, Chesterfield Royal Hospital Calow, Chesterfield, England, UK.
  • Coward TJ; Professor and Honorary Consultant in Maxillofacial and Craniofacial Rehabilitation, Academic Centre of Reconstructive Science, Faculty of Dentistry, Oral and Craniofacial Sciences, King's College London, London, England, UK.
  • Shuweihdi F; Lecturer in Medical Statistics and Health Data Science, Dental Translational and Clinical Research Unit, School of Dentistry, University of Leeds, Leeds, England, UK.
  • Bojke C; Professor of Health Economics, Academic Unit of Health Economics, Leeds Institute of Health Sciences, University of Leeds, Leeds, England, UK.
  • Pavitt SH; Professor of Translational and Applied Health Research, Dental Translational and Clinical Research Unit, School of Dentistry, University of Leeds, Leeds, England, UK.
  • Nattress BR; Emeritus Professor of Restorative Dentistry, Department of Restorative Dentistry, School of Dentistry, University of Leeds, Leeds, England, UK.
  • Keeling AJ; Professor of Prosthodontics and Digital Dentistry, Department of Restorative Dentistry, School of Dentistry, University of Leeds, Leeds, England, UK.
J Prosthet Dent ; 2024 Apr 13.
Article in En | MEDLINE | ID: mdl-38616155
ABSTRACT
STATEMENT OF

PROBLEM:

Manually sculpting a wax pattern of a facial prosthesis is a time-, skill-, and resource-intensive process. Computer-aided design (CAD) methods have been proposed as a substitute for manual sculpting, but these techniques can still require high technical or artistic abilities. Three-dimensional morphable models (3DMMs) could semi-automate facial prosthesis CAD. Systematic comparisons of different design approaches are needed.

PURPOSE:

The purpose of this study was to compare the trueness and repeatability of replacing facial features with 3 methods of facial prosthesis design involving 3DMM, traditional CAD, and conventional manual sculpting techniques. MATERIAL AND

METHODS:

Fifteen participants without facial defects were scanned with a structured light scanner. The facial meshes were manipulated to generate artificial orbital, nasal, or combined defects. Three methods of facial prosthesis design were compared for the 15 participants and repeated to produce 5 of each design for 2 participants. For the 3DMM approach, the Leeds face model informed the designs in a statistically meaningful way. For the traditional CAD methods, designs were created by using mirroring techniques or from a nose model database. For the conventional manual sculpting techniques, wax patterns were manually created on 3D printed full face baseplates. For analysis, the unedited facial feature was the standard. The unsigned distance was calculated from each of the several thousand vertices on the unedited facial feature to the closest point on the external surface of the prosthesis prototype. The mean absolute error was calculated, and a Friedman test was performed (α=.05).

RESULTS:

The median mean absolute error was 1.13 mm for the 3DMM group, 1.54 mm for the traditional CAD group, and 1.49 mm for the manual sculpting group, with no statistically significant differences among groups (P=.549). Boxplots showed substantial differences in the distribution of mean absolute error among groups, with the 3DMM group showing the greatest consistency. The 3DMM approach produced repeat designs with the lowest coefficient of variation.

CONCLUSIONS:

The 3DMM approach shows potential as a semi-automated method of CAD. Further clinical research is planned to explore the 3DMM approach in a feasibility trial.

Full text: 1 Collection: 01-internacional Database: MEDLINE Language: En Journal: J Prosthet Dent Year: 2024 Document type: Article