A simple light path modifying device to reduce scattering in front-face fluorescence spectra.
Peters, Frank B; Rapp, Andreas O.
Affiliation
  • Peters FB; Institut für Berufswissenschaften im Bauwesen, Leibniz-Universität Hannover, Herrenhäuser Straße 8, 30419 Hannover, Germany.
  • Rapp AO; Institut für Berufswissenschaften im Bauwesen, Leibniz-Universität Hannover, Herrenhäuser Straße 8, 30419 Hannover, Germany.
Methods Appl Fluoresc ; 12(3)2024 Jun 17.
Article in En | MEDLINE | ID: mdl-38834089
ABSTRACT
This technical note presents a device to diminish scattering signal in front-face fluorescence spectra while obtaining fluorescence signal. The beam path in a commercial fluorescence spectrometer was modified by two deflecting mirrors, leading reflections away from the sensor. This light path modifying (LPM) device was tested with two fluid and three solid substances, where the scattering-to-fluorescence ratio improved by a factor of 1.7 to 7.6. The spectra obtained with the LPM were much clearer, and distortion of the fluorescence peaks was avoided. Scans of quinine sulphate complied well with reference spectra.