Your browser doesn't support javascript.
loading
Accelerated B 1 + $$ {\mathrm{B}}_1
Egger, Nico; Nagelstraßer, Sophia; Wildenberg, Saskia; Bitz, Andreas; Ruck, Laurent; Herrler, Jürgen; Meixner, Christian Rainer; Kimmlingen, Ralph; Lanz, Titus; Schmitter, Sebastian; Uder, Michael; Nagel, Armin Michael.
Affiliation
  • Egger N; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Nagelstraßer S; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Wildenberg S; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Bitz A; Electrical Engineering and Information Technology, University of Applied Sciences - FH Aachen, Aachen, Germany.
  • Ruck L; Electrical Engineering and Information Technology, University of Applied Sciences - FH Aachen, Aachen, Germany.
  • Herrler J; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
  • Meixner CR; Siemens Healthcare, Erlangen, Bavaria, Germany.
  • Kimmlingen R; Siemens Healthcare, Erlangen, Bavaria, Germany.
  • Lanz T; Siemens Healthcare, Erlangen, Bavaria, Germany.
  • Schmitter S; Rapid Biomedical, Rimpar, Germany.
  • Uder M; Physikalisch-Technische Bundesanstalt (PTB), Braunschweig and Berlin, Germany.
  • Nagel AM; Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
Magn Reson Med ; 92(5): 1933-1951, 2024 Nov.
Article in En | MEDLINE | ID: mdl-38888143
ABSTRACT

PURPOSE:

To investigate the impact of reduced k-space sampling on B 1 + $$ {\mathrm{B}}_1^{+} $$ mapping and the resulting impact on phase shimming and dynamic/universal parallel transmit (pTx) RF pulse design.

METHODS:

Channel-wise 3D B 1 + $$ {\mathrm{B}}_1^{+} $$ maps were measured at 7 T in 35 and 23 healthy subjects for the heart and prostate region, respectively. With these B 1 + $$ {\mathrm{B}}_1^{+} $$ maps, universal phase shims optimizing homogeneity and B 1 + $$ {\mathrm{B}}_1^{+} $$ efficiency were designed for heart and prostate imaging. In addition, universal 4kT-point pulses were designed for the heart. Subsequently, individual phase shims and individual 4kT-pulses were designed based on B 1 + $$ {\mathrm{B}}_1^{+} $$ maps with different acceleration factors and tested on the original maps. The performance of the pulses was compared by evaluating their coefficients of variation (CoV), B 1 + $$ {\mathrm{B}}_1^{+} $$ efficiencies and specific energy doses (SED). Furthermore, validation measurements were carried out for one heart and one prostate subject.

RESULTS:

For both organs, the universal phase shims showed significantly higher B 1 + $$ {\mathrm{B}}_1^{+} $$ efficiencies and lower CoVs compared to the vendor provided default shim, but could still be improved with individual phase shims based on accelerated B 1 + $$ {\mathrm{B}}_1^{+} $$ maps (acquisition time = 30 s). In the heart, the universal 4kT-pulse achieved significantly lower CoVs than tailored phase shims. Tailored 4kT-pulses based on accelerated B 1 + $$ {\mathrm{B}}_1^{+} $$ maps resulted in even further reduced CoVs or a 2.5-fold reduction in SED at the same CoVs as the universal 4kT-pulse.

CONCLUSION:

Accelerated B 1 + $$ {\mathrm{B}}_1^{+} $$ maps can be used for the design of tailored pTx pulses for prostate and cardiac imaging at 7 T, which further improve homogeneity, B 1 + $$ {\mathrm{B}}_1^{+} $$ efficiency, or SED compared to universal pulses.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Prostate / Algorithms / Magnetic Resonance Imaging / Heart Limits: Adult / Humans / Male Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2024 Document type: Article Affiliation country: Country of publication:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Prostate / Algorithms / Magnetic Resonance Imaging / Heart Limits: Adult / Humans / Male Language: En Journal: Magn Reson Med Journal subject: DIAGNOSTICO POR IMAGEM Year: 2024 Document type: Article Affiliation country: Country of publication: