Your browser doesn't support javascript.
loading
Comparison of the repeatability and reproducibility of corneal thickness mapping using optical coherence tomography according to tear film break-up time.
Lin, Kan; Xu, Zhiqiang; Wang, Hui; Wang, Yuzhou; Wei, Linzhi; Ma, Hongqing; Zhao, Jian; Lu, Fan; Hu, Liang.
Affiliation
  • Lin K; National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, China.
  • Xu Z; National Engineering Research Center of Ophthalmology and Optometry, Eye Hospital, Wenzhou Medical University, Wenzhou, 325027, China.
  • Wang H; Zhejiang University School of Medicine Second Affiliated Hospital Eye Center, 88 Jiefang Road, Hangzhou, 310009, China.
  • Wang Y; National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, China.
  • Wei L; National Engineering Research Center of Ophthalmology and Optometry, Eye Hospital, Wenzhou Medical University, Wenzhou, 325027, China.
  • Ma H; National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, China.
  • Zhao J; National Engineering Research Center of Ophthalmology and Optometry, Eye Hospital, Wenzhou Medical University, Wenzhou, 325027, China.
  • Lu F; National Clinical Research Center for Ocular Diseases, Eye Hospital, Wenzhou Medical University, Wenzhou, Zhejiang, China.
  • Hu L; National Engineering Research Center of Ophthalmology and Optometry, Eye Hospital, Wenzhou Medical University, Wenzhou, 325027, China.
BMC Ophthalmol ; 24(1): 275, 2024 Jul 05.
Article in En | MEDLINE | ID: mdl-38970043
ABSTRACT

BACKGROUND:

To compare the repeatability and reproducibility of corneal and corneal epithelial thickness mapping using anterior segment optical coherence tomography (AS-OCT) according to tear film break-up time (TBUT).

METHODS:

The included eyes were divided into three subgroups according to TBUT (group 1 TBUT ≤ 5 s, group 2 5 s < TBUT ≤ 10 s, and group 3 TBUT > 10 s). All eyes were imaged separately thrice by two operators to obtain the thickness maps (TMs) of the cornea and corneal epithelium based on spatial zones encompassing a 9-mm-diameter area. Each TM consisted of 25 areas. Intraoperator (repeatability) and interoperator (reproducibility) standard deviations (Sws), coefficients of variation (CoVs), and intraclass correlation coefficients (ICCs) among the tests were calculated and compared in all the areas.

RESULTS:

Altogether, 132 eyes of 67 subjects were included (50, 47, and 35 eyes in groups 1, 2, and 3; respectively). The ICCs of corneal epithelial thickness and corneal thickness were > 0.75 in most of the areas. Pairwise comparisons showed that AS-OCT exhibited lower repeatability in group 1 than in groups 2 and 3 (P < 0.05). However groups 2 and 3 showed similar results. Sws and CoVs of corneal epithelial thickness exhibited no significant interoperator differences. While no significant differences were observed in corneal thickness in most of the areas.

CONCLUSIONS:

TBUT significantly influences the repeatability of corneal and corneal epithelial thickness measurements. Poor tear film stability requires careful evaluation of corneal epithelial thickness.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Tears / Cornea / Tomography, Optical Coherence Limits: Adult / Aged / Female / Humans / Male / Middle aged Language: En Journal: BMC Ophthalmol Journal subject: OFTALMOLOGIA Year: 2024 Document type: Article Affiliation country: Country of publication:

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Tears / Cornea / Tomography, Optical Coherence Limits: Adult / Aged / Female / Humans / Male / Middle aged Language: En Journal: BMC Ophthalmol Journal subject: OFTALMOLOGIA Year: 2024 Document type: Article Affiliation country: Country of publication: