Your browser doesn't support javascript.
loading
Comparison of in vivo and in vitro models to evaluate pulp temperature rise during exposure to a Polywave® LED light curing unit
Runnacles, Patricio; Arrais, Cesar Augusto Galvão; Maucoski, Cristiane; Coelho, Ulisses; De Goes, Mario Fernando; Rueggeberg, Frederick Allen.
Afiliação
  • Runnacles, Patricio; Universidade Estadual de Ponta Grossa. Departamento de Odontologia Restauradora. Paraná. BR
  • Arrais, Cesar Augusto Galvão; Universidade Estadual de Ponta Grossa. Departamento de Odontologia Restauradora. Paraná. BR
  • Maucoski, Cristiane; Universidade Estadual de Ponta Grossa. Departamento de Odontologia Restauradora. Paraná. BR
  • Coelho, Ulisses; Universidade Estadual de Ponta Grossa. Departamento de Odontologia Restauradora. Paraná. BR
  • De Goes, Mario Fernando; Universidade Estadual de Campinas. Faculdade de Odontologia de Piracicaba. Departamento de Materiais Dentários. São Paulo. BR
  • Rueggeberg, Frederick Allen; Augusta University. Dental College of Georgia. Department of Restorative Sciences. Georgia. US
J. appl. oral sci ; 27: e20180480, 2019. tab, graf
Artigo em Inglês | LILACS, BBO - Odontologia | ID: biblio-1002402
Biblioteca responsável: BR1.1
ABSTRACT
Abstract

Objectives:

To measure and compare in vivo and in vitro pulp temperature (PT) increase (ΔTEMP) over baseline, physiologic temperature using the same intact upper premolars exposed to the same Polywave® LED curing light.

Methodology:

After local Ethics Committee approval (#255,945), local anesthesia, rubber dam isolation, small occlusal preparations/minute pulp exposure (n=15) were performed in teeth requiring extraction for orthodontic reasons. A sterile probe of a temperature measurement system (Temperature Data Acquisition, Physitemp) was placed within the pulp chamber and the buccal surface was sequentially exposed to a LED LCU (Bluephase 20i, Ivoclar Vivadent) using the following exposure modes 10-s low or high, 5-s Turbo, and 60-s high. Afterwards, the teeth were extracted and K-type thermocouples were placed within the pulp chamber through the original access. The teeth were attached to an assembly simulating the in vivo environment, being similarly exposed while real-time temperature (°C) was recorded. ΔTEMP values and time for temperature to reach maximum (ΔTIME) were subjected to two-way ANOVA and Bonferroni's post-hoc tests (pre-set alpha 0.05).

Results:

Higher ΔTEMP was observed in vitro than in vivo. No significant difference in ΔTIME was observed between test conditions. A significant, positive relationship was observed between radiant exposure and ΔTEMP for both conditions (in vivo r2=0.917; p<0.001; in vitro r2=0.919; p<0.001).

Conclusion:

Although the in vitro model overestimated in vivo PT increase, in vitro PT rise was close to in vivo values for clinically relevant exposure modes.
Assuntos


Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: BBO - Odontologia / LILACS Assunto principal: Temperatura / Polpa Dentária / Luzes de Cura Dentária Tipo de estudo: Estudo diagnóstico / Estudo prognóstico Aspecto: Aspectos éticos Limite: Humanos Idioma: Inglês Revista: J. appl. oral sci Assunto da revista: Odontologia Ano de publicação: 2019 Tipo de documento: Artigo País de afiliação: Brasil / Estados Unidos Instituição/País de afiliação: Augusta University/US / Universidade Estadual de Campinas/BR / Universidade Estadual de Ponta Grossa/BR

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: BBO - Odontologia / LILACS Assunto principal: Temperatura / Polpa Dentária / Luzes de Cura Dentária Tipo de estudo: Estudo diagnóstico / Estudo prognóstico Aspecto: Aspectos éticos Limite: Humanos Idioma: Inglês Revista: J. appl. oral sci Assunto da revista: Odontologia Ano de publicação: 2019 Tipo de documento: Artigo País de afiliação: Brasil / Estados Unidos Instituição/País de afiliação: Augusta University/US / Universidade Estadual de Campinas/BR / Universidade Estadual de Ponta Grossa/BR
...