Your browser doesn't support javascript.
loading
A strategy for designing of customized electromechanical actuators of blood pumps
Souza, Rogerio Lima de; Chabu, Ivan Eduardo; Silva, Evandro Drigo da; Andrade, Aron Jose Pazin de; Leao, Tarcisio Fernandes; Bock, Eduardo Guy Perpetuo.
Afiliação
  • Souza, Rogerio Lima de; Federal Institute of Technology in Sao Paulo IFSP. São Paulo. BR
  • Chabu, Ivan Eduardo; University of São Paulo. São Paulo. BR
  • Silva, Evandro Drigo da; Federal Institute of Technology in Sao Paulo IFSP. Institute Dante Pazzanese of Cardiology IDPC. São Paulo. BR
  • Andrade, Aron Jose Pazin de; Institute Dante Pazzanese of Cardiology IDPC. São Paulo. BR
  • Leao, Tarcisio Fernandes; Federal Institute of Technology in Sao Paulo IFSP. Institute Dante Pazzanese of Cardiology IDPC. São Paulo. BR
  • Bock, Eduardo Guy Perpetuo; Institute of Technology in Sao Paulo IFSP. Institute Dante Pazzanese of Cardiology IDPC. São Paulo. BR
Artif. organs ; 44(8): 797-802, Aug. 2020. ilus., graf.
Artigo em Inglês | Sec. Est. Saúde SP, SESSP-IDPCPROD, Sec. Est. Saúde SP | ID: biblio-1121435
Biblioteca responsável: BR79.1
ABSTRACT
Congestive heart failure is a pathology of global incidence that affects millions of people worldwide. When the heart weakens and fails to pump blood at physiological rates commensurate with the requirements of tissues, two main alternatives are cardiac transplant and ventricular assist devices (VADs). This article presents the design strategy for development of a customized VAD electromagnetic actuator. Electromagnetic actuator is a brushless direct current motor customized to drive the pump impeller by permanent magnets located in rotor­stator coupling. In this case, ceramic pivot bearings support the VAD impeller. Electronic circuitry controls rotation switching current in stator coils. The proposed methodology consisted of analytical numerical design, tridimensional computational modeling, numerical simulations using Maxwell software, actuator prototyping, and validation in the dynamometer. The axial flow actuator was chosen by its size and high power density compared to the radial flow type. First step consisted of estimating the required torque to drive the pump. Torque was estimated at 2100 rpm and mean current of 0.5 A. Numerical analysis using finite element method mapped vectors and fields to build stator coils and actuator assemblage. After tests in the dynamometer, experimental results were compared with numerical simulation and validated the proposed model. In conclusion, the proposed methodology for designing of VAD electromechanical actuator was considered satisfactory in terms of data consistency, feasibility, and reliability.
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Brasil Base de dados: Sec. Est. Saúde SP / SESSP-IDPCPROD Assunto principal: Materiais Biocompatíveis / Coração Auxiliar / Desenho de Equipamento Idioma: Inglês Revista: Artif. organs Ano de publicação: 2020 Tipo de documento: Artigo Instituição/País de afiliação: Federal Institute of Technology in Sao Paulo IFSP/BR / Institute Dante Pazzanese of Cardiology IDPC/BR / University of São Paulo/BR

Texto completo: Disponível Coleções: Bases de dados nacionais / Brasil Base de dados: Sec. Est. Saúde SP / SESSP-IDPCPROD Assunto principal: Materiais Biocompatíveis / Coração Auxiliar / Desenho de Equipamento Idioma: Inglês Revista: Artif. organs Ano de publicação: 2020 Tipo de documento: Artigo Instituição/País de afiliação: Federal Institute of Technology in Sao Paulo IFSP/BR / Institute Dante Pazzanese of Cardiology IDPC/BR / University of São Paulo/BR
...