Your browser doesn't support javascript.
loading
Determinación niveles de glucosa por medio de espectroscopía como método no invasivo / Determination of glucose levels by means of spectroscopy as a non-invasive method
Briseño-Castellanos, Montserrat; Hernández-Gonzalez, Martha Alicia; Ramos-Moreno, José Miguel; Cisneros-Carrasco, Juan Manuel; Jiménez-Ruvalcaba, Jabel.
Afiliação
  • Briseño-Castellanos, Montserrat; Instituto Mexicano del Seguro Social. Centro Médico Nacional del Bajío, Hospital de Especialidades No. 1. Servicio de Medicina Interna. León, Guanajuato. MX
  • Hernández-Gonzalez, Martha Alicia; Instituto Mexicano del Seguro Social. Centro Médico Nacional del Bajío, Hospital de Especialidades No. 1. División de la Investigación en Salud. León, Guanajuato. MX
  • Ramos-Moreno, José Miguel; Universidad de Guanajuato Campus León. División de Ciencias e Ingenierías. León, Guanajuato. MX
  • Cisneros-Carrasco, Juan Manuel; Instituto Mexicano del Seguro Social. Centro Médico Nacional del Bajío, Hospital de Especialidades No. 1. Laboratorio Clínico. León, Guanajuato. MX
  • Jiménez-Ruvalcaba, Jabel; Universidad de Guanajuato Campus León. División de Ciencias e Ingenierías. León, Guanajuato. MX
Rev. Méd. Inst. Mex. Seguro Soc ; 59(6): 517-527, dic. 2021. ilus, tab, graf
Artigo em Espanhol | LILACS | ID: biblio-1357536
Biblioteca responsável: MX346.1
RESUMEN

Introducción:

la diabetes mellitus es uno de los problemas más graves de salud pública que enfrenta México. El factor más preocupante es la falta de control de la misma, lo que incide de manera directa, causando daños severos a la salud y la calidad de vida del paciente y familiares, así como una carga económica al país. Por lo tanto, el desarrollo de un método no invasivo para la medición de la glucemia proporcionaría a los pacientes una forma sencilla e indolora de monitoreo y, en consecuencia, un mejor control de la diabetes.

Objetivo:

investigar, desarrollar y validar un sensor no invasivo por medio de la espectroscopía para la estimación del nivel de glucosa en sangre. Material y

métodos:

se realizó un análisis de estudio transversal analítico de correlación realizado en las instalaciones del laboratorio de la UMAE No 1, Bajío. Se incluyeron pacientes adultos voluntarios que acudieron al laboratorio de dicha unidad para la toma de niveles de glucosa sérica y de manera simultánea se realizó la medición a través de método no invasivo por espectroscopía y, posteriormente, se compararon ambos resultados para demostrar la validez del dispositivo.

Resultados:

mediante el análisis de la diferencia de medias de Bland-Altman, se identificó que solamente un paciente tuvo un valor extremo, y que el método para medir la glucosa de manera no invasiva sobreestima hasta un 10.2% del valor de glucosa central.

Conclusión:

comparando dichos resultados con las normas para glucómetros digitales se concluye que nuestro dispositivo es capaz de proporcionar niveles de glucosa certeros.
ABSTRACT

Background:

Diabetes mellitus is one of the most serious public health problems in Mexico. The most worrying factor is the lack of control of it, which has a direct impact, causing severe damage to the health and quality of life of the patient and its family, as well as an economic burden to the health system. Therefore, the development of a non-invasive method for measuring blood glucose would provide to patients a simple and painless way of monitoring and consequently better control of diabetes.

Objective:

Research, development and validation of a non-invasive sensor by means of spectroscopy for the estimation of the blood glucose level. Material and

methods:

An analysis of a cross-sectional analytical correlation study was carried out in the facilities of the laboratory at the UMAE No. 1, Bajío. Voluntary adult patients who attended the laboratory of the UMAE to take serum glucose levels were included, and simultaneously the measurement was carried out through a non-invasive method by spectroscopy and, later, both results were compared to demonstrate the validity of the device.

Results:

By the Bland-Altman mean difference analysis, it was identified that only one patient had an extreme value, and that the method to measure glucose non-invasively overestimates up to 10.2% of the central glucose value.

Conclusion:

Comparing these results with the standards for digital glucometers, it is concluded that our device is capable of providing accurate glucose levels.
Licença
Assuntos

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Análise Espectral / Glicemia / Diabetes Mellitus / México Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo de prevalência / Estudo prognóstico Aspecto: Preferência do paciente Limite: Feminino / Humanos / Masculino País/Região como assunto: México Idioma: Espanhol Revista: Rev. Méd. Inst. Mex. Seguro Soc Assunto da revista: Medicina / Medicina Social Ano de publicação: 2021 Tipo de documento: Artigo País de afiliação: México Instituição/País de afiliação: Instituto Mexicano del Seguro Social/MX / Universidad de Guanajuato Campus León/MX
Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Análise Espectral / Glicemia / Diabetes Mellitus / México Tipo de estudo: Estudo diagnóstico / Estudo observacional / Estudo de prevalência / Estudo prognóstico Aspecto: Preferência do paciente Limite: Feminino / Humanos / Masculino País/Região como assunto: México Idioma: Espanhol Revista: Rev. Méd. Inst. Mex. Seguro Soc Assunto da revista: Medicina / Medicina Social Ano de publicação: 2021 Tipo de documento: Artigo País de afiliação: México Instituição/País de afiliação: Instituto Mexicano del Seguro Social/MX / Universidad de Guanajuato Campus León/MX
...