Your browser doesn't support javascript.
loading
Identificação radiológica de dispositivos cardíacos eletrônicos implantáveis / X-ray identification of implantable electronic cardiac rhythm management devices
Valdigem, Bruno Pereira; Silva, Nilton José Carneiro da; Andalaft, Rogério Braga; Melo, Celso Salgado de.
Afiliação
  • Valdigem, Bruno Pereira; Instituto Dante Pazzanese de Cardiologia. Hospital Israelita Albert Einstein. São Paulo - SP. BR
  • Silva, Nilton José Carneiro da; Hospital Israelita Albert Einstein. Eletrofisiologista. São Paulo - SP. BR
  • Andalaft, Rogério Braga; Hospital Israelita Albert Einstein. Centro de Arritmias Cardíacas. Cardiologia Pediátrica e Arritmologista. São Paulo - SP. BR
  • Melo, Celso Salgado de; Universidade Federal do Triângulo Mineiro. Serviço de Marcapassos. Eletrofisiologia e Estimulação Cardíaca Artificial. Uberaba - MG. BR
RELAMPA, Rev. Lat.-Am. Marcapasso Arritm ; 30(1): 3-8, jan.-mar. 2017. ilus, graf
Artigo em Português | LILACS, Sec. Est. Saúde SP, SESSP-IDPCPROD, Sec. Est. Saúde SP | ID: biblio-837227
Biblioteca responsável: BR44.1
Localização: BR79.1
RESUMO

Introdução:

Este trabalho teve por objetivos identificar as características comuns aos geradores e classificá-los no menor número possível de grupos para facilitar a identificação radiológica, e, também, identificar características que permitam diferenciar geradores DF1 de DF4.

Método:

A amostragem foi composta por 62 imagens de geradores de dispositivos implantáveis bem como de cabos-eletrodos comercialmente disponíveis no Brasil de 2008 até o presente momento das empresas Biotronik, Medtronic, St. Jude Medical e Boston Scientific. As imagens consideradas com qualidade adequada de cada modelo de dispositivo implantável foram selecionadas em busca de características comuns a cada fabricante.

Resultados:

A revisão criteriosa das características de cada fabricante e modelo possibilitou divisar claramente oito categorias, com características únicas de cada fabricante para estimuladores (grupo 1) e cardiodesfibriladores (grupo 2). Um algoritmo de identificação foi desenvolvido, a partir das características mais facilmente identificáveis de cada grupo.

Conclusão:

A identificação radiológica dos dispositivos implantáveis é possível e confiável. Os cabos-eletrodos de mesma função (choque ou estimulação) não puderam ser diferenciados em relação ao fabricante, mas algumas características das conexões de cabos-eletrodos de choque podem diferenciar a conexão DF1 de DF4, bem como mono coil de duplo coil
ABSTRACT

Background:

This study is aimed at identifying common characteristics of generators and classify them into the lowest possible number of groups to facilitate X-ray identification as well as to identify characteristics to differentiate DF1 and DF4 generators.

Method:

The sample included 62 images of implantable device generators and leads available in the Brazilian market from 2008 to the present moment manufactured by Biotronik, Medtronic, St. Jude Medical and Boston Scientific. Good quality images from each implantable device model were chosen to help identify common characteristics of each manufacturer.

Results:

A careful review of manufacturer's characteristics and model enabled a clear division into eight categories with each manufacturer's unique characteristics for pacers (group 1) and cardiac defibrillators (group 2). An identification algorhythm was developed based on the most easily identifiable characteristics of each group.

Conclusion:

X-ray identification of implantable devices is possible and reliable. Leads with the same function (shock or pacing) were not differentiated by manufacturer, but some shock lead connection characteristics may differentiate DF1 from DF4 connections, as well monocoil from double coil
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Brasil Base de dados: LILACS / Sec. Est. Saúde SP / SESSP-IDPCPROD Assunto principal: Raios X / Radiografia / Desfibriladores Implantáveis Tipo de estudo: Estudo diagnóstico / Estudo prognóstico Limite: Humanos Idioma: Português Revista: RELAMPA, Rev. Lat.-Am. Marcapasso Arritm Ano de publicação: 2017 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Israelita Albert Einstein/BR / Instituto Dante Pazzanese de Cardiologia/BR / Universidade Federal do Triângulo Mineiro/BR

Texto completo: Disponível Coleções: Bases de dados nacionais / Brasil Base de dados: LILACS / Sec. Est. Saúde SP / SESSP-IDPCPROD Assunto principal: Raios X / Radiografia / Desfibriladores Implantáveis Tipo de estudo: Estudo diagnóstico / Estudo prognóstico Limite: Humanos Idioma: Português Revista: RELAMPA, Rev. Lat.-Am. Marcapasso Arritm Ano de publicação: 2017 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Israelita Albert Einstein/BR / Instituto Dante Pazzanese de Cardiologia/BR / Universidade Federal do Triângulo Mineiro/BR
...