Tomografía computarizada de doble energía: nueva tecnología para la reducción de artefactos de metal / Dual-energy computed tomography: new technology for metal artifacts reduction
Reynoso, Exequiel; Carrascosa, Patricia; Capuñay, Carlos; Rasumoff, Alejandro; Vallejos, Javier; Carpio, Jimena; Lago, Karen.
Afiliação
  • Reynoso, Exequiel; Diagnóstico Maipú. Buenos Aires. AR
  • Carrascosa, Patricia; Diagnóstico Maipú. Buenos Aires. AR
  • Capuñay, Carlos; Diagnóstico Maipú. Buenos Aires. AR
  • Rasumoff, Alejandro; Diagnóstico Maipú. Buenos Aires. AR
  • Vallejos, Javier; Diagnóstico Maipú. Buenos Aires. AR
  • Carpio, Jimena; Diagnóstico Maipú. Buenos Aires. AR
  • Lago, Karen; Diagnóstico Maipú. Buenos Aires. AR
Rev. Asoc. Argent. Ortop. Traumatol ; 82(1): 5-13, mar. 2017. []
Artigo em Espanhol | LILACS, BINACIS | ID: biblio-842504
Biblioteca responsável: AR337.1
RESUMEN

Introducción:

El objetivo fue explorar la utilidad de la tomografía computarizada de doble energía mediante tecnología de imágenes espectrales gemstone y de un programa destinado a la reducción de artefactos de metal (MARS), para evaluar tejidos periprotésicos, y la interpretabilidad diagnóstica de patologías relacionadas con implantes. Materiales y

Métodos:

Se comparó la densidad ósea, de partes blandas y de grasa en el tejido periprotésico y en tejido de control sin implante, utilizando un escáner de alta definición de tomografía computarizada de doble energía tanto en imágenes policromáticas convencionales, como en monocromáticas virtuales con MARS, en 80 pacientes con prótesis metálicas en diversas regiones musculoesqueléticas. Se valoró la calidad de imagen y la interpretabilidad diagnóstica mediante la escala de Likert.

Resultados:

Con imágenes policromáticas hubo diferencias significativas entre el área periprotésica en los tres tejidos respecto a los controles (p <0,0001); sin diferencias significativas utilizando imágenes espectrales monocromáticas virtuales-MARS (hueso p = 0,053, partes blandas p = 0,32 y grasa p = 0,13), con más similitud con el tejido normal. Los niveles de ruido fueron significativamente mayores con imágenes policromáticas (p <0,0001) que con imágenes espectrales monocromáticas virtuales-MARS. Se consideraron no interpretables todas las regiones periprotésicas en las imágenes policromáticas y 11 (9%) en las imágenes espectrales monocromáticas virtuales-MARS. No hubo diferencias significativas en la dosis de radiación comparada con la del grupo control (p = 0,21).

Conclusiones:

La tomografía computarizada de doble energía puede reducir los artefactos periprotésicos, logrando un significativo incremento en la capacidad de identificar tejidos y la interpretabilidad diagnóstica de posibles patologías relacionadas con implantes. Nivel de Evidencia II
ABSTRACT

Introduction:

To explore the usefulness of dual energy imaging using gemstone spectral imaging technology and a dedicated software for metal artifact reduction (MARS) for the evaluation of periprosthetic tissues, and to assess image interpretability of implant-related complications.

Methods:

Signal density measurements were performed in periprosthetic and remote (control) areas in bone, soft tissue, and fat among 80 patients using a high definition scanner. Polychromatic images and virtual monochromatic spectral images with MARS were obtained, and image quality and diagnostic interpretability were evaluated using a Likert scale.

Results:

Using polychromatic images, the periprosthetic area showed significant differences compared to the remote areas among the three tissue explored (p<0.0001 for all); with no significant differences using virtual monochromatic spectral images-MARS (bone p=0.053, soft tissue p=0.32, fat p=0.13), suggesting similar signal density compared to normal (remote) tissue. Furthermore, periprosthetic polychromatic image noise levels were significantly higher than with virtual monochromatic spectral images-MARS (p<0.0001). All periprosthetic areas were deemed non-interpretable using polychromatic images, compared to 11 (9%) using virtual monochromatic spectral images-MARS. There were no differences in radiation dose compared to control group (p=0.21).

Conclusions:

Virtual monochromatic spectral images-MARS technology has the ability to reduce periprosthetic artifacts, achieving a significant increase to identify tissues and diagnostic interpretability of complications related to the implants. Level of Evidence II
Assuntos


Texto completo: Disponível Coleções: Bases de dados internacionais Contexto em Saúde: Agenda de Saúde Sustentável para as Américas Problema de saúde: Objetivo 5 Medicamentos, vacinas e tecnologias sanitárias Base de dados: BINACIS / LILACS Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Intensificação de Imagem Radiográfica / Tomografia Computadorizada por Raios X / Artefatos / Prótese Articular / Metais Tipo de estudo: Estudo observacional Idioma: Espanhol Revista: Rev. Asoc. Argent. Ortop. Traumatol Assunto da revista: Ortopedia Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: Argentina Instituição/País de afiliação: Diagnóstico Maipú/AR

Texto completo: Disponível Coleções: Bases de dados internacionais Contexto em Saúde: Agenda de Saúde Sustentável para as Américas Problema de saúde: Objetivo 5 Medicamentos, vacinas e tecnologias sanitárias Base de dados: BINACIS / LILACS Assunto principal: Interpretação de Imagem Radiográfica Assistida por Computador / Intensificação de Imagem Radiográfica / Tomografia Computadorizada por Raios X / Artefatos / Prótese Articular / Metais Tipo de estudo: Estudo observacional Idioma: Espanhol Revista: Rev. Asoc. Argent. Ortop. Traumatol Assunto da revista: Ortopedia Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: Argentina Instituição/País de afiliação: Diagnóstico Maipú/AR
...