Avaliação in vivo do torque de inserção do novo sistema de implantes Emfils / In vivo insertion torque assessment of the new dental implant system Emfi ls
Campos, Thiago Barros; Quinteiro, Ana Paula Moro; Bassi, Ana Paula Farnezi; Ponzoni, Daniela; Carvalho, Paulo Sérgio Perri de.
Afiliação
  • Campos, Thiago Barros; s.af
  • Quinteiro, Ana Paula Moro; s.af
  • Bassi, Ana Paula Farnezi; s.af
  • Ponzoni, Daniela; Unesp. Faculdade de Odontologia de Araçatuba. Depto. de Cirurgia e Clínica Integrada. Araçatuba. BR
  • Carvalho, Paulo Sérgio Perri de; Unesp. Faculdade de Odontologia de Araçatuba. Depto. de Cirurgia e Clínica Integrada. Araçatuba. BR
ImplantNewsPerio ; 2(3): 425-432, mai.-jun. 2017. ilus, tab
Artigo em Português | LILACS, BBO - Odontologia | ID: biblio-847248
Biblioteca responsável: BR510.1
RESUMO

Objetivo:

avaliar o torque de inserção de um novo desenho de implante. Material e

métodos:

pacientes foram recrutados para instalação de implantes, no período de março a dezembro de 2015. Dados como o diâmetro, o comprimento do implante, a sequência de fresagem, a área de instalação e o torque atingido foram considerados. A avaliação do torque foi realizada com um torquímetro cirúrgico.

Resultados:

foram avaliados 82 implantes, com diâmetros entre 3,5 mm e 5 mm, e comprimentos entre 7 mm e 16 mm, sendo instalados 46 implantes na mandíbula e 36 implantes na maxila. Os torques (média ± desvio-padrão) alcançados foram de 53 ± 9,9 Ncm para maxila e 57 ± 11,3 Ncm para mandíbula.

Conclusão:

o sistema de implantes Novo Colosso promoveu torque de inserção adequado, com boa estabilidade primária nos implantes.
ABSTRACT

Objective:

to evaluate the insertion torque of a new implant design. Material and

methods:

patients were recruited for implant placement from March to December 2015. Data such as diameter, implant length, drilling sequence, installation area, and torque reached were considered. Torque evaluation was performed with a surgical torque wrench.

Results:

82 implants were evaluated, with diameters between 3.5 mm and 5.0 mm, and lengths between 7 mm and 16 mm, with 46 implants in the mandible and 36 implants in the maxilla. The torques (mean ± standard deviation) reached 53 ± 9.9 Ncm for maxilla and 57 ± 11.3 Ncm for mandible.

Conclusion:

the Novo Colosso implant system promotes adequate insertion torque with good primary implant stability.
Assuntos

Buscar no Google
Coleções: Bases de dados internacionais Base de dados: BBO - Odontologia / LILACS Assunto principal: Osteotomia / Implantes Dentários / Osseointegração / Torque Tipo de estudo: Estudo de avaliação Limite: Humanos Idioma: Português Revista: ImplantNewsPerio Assunto da revista: Odontologia Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Unesp/BR
Buscar no Google
Coleções: Bases de dados internacionais Base de dados: BBO - Odontologia / LILACS Assunto principal: Osteotomia / Implantes Dentários / Osseointegração / Torque Tipo de estudo: Estudo de avaliação Limite: Humanos Idioma: Português Revista: ImplantNewsPerio Assunto da revista: Odontologia Ano de publicação: 2017 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Unesp/BR
...