Abstract Objective:
To report the results of four mock femoral venous cannulas and the hydrodynamical superiority of one of them, which is the completely punched (CP) model, upon the other three.Methods:
Four simulated femoral venous cannulas (single-stage, two-stage, multi-stage, and CP model) were designed from a 1/4" x 1/16" x 68 cm polyvinyl chloride (PVC) tubing line for testing. Holes on the PVC tubes were opened by a 5 mm aortic punch. In order to evaluate the cannulas' drainage performance, gelofusine was used as fluid. The fluid was drained for 60 seconds by gravitation and then measured for each model separately.Results:
Mean drained volumes of single-stage, two-stage, and multi-stage cannulas were 2.483, 2.561, and 2.603 mL, respectively. However, the CP cannula provided us a mean drained volume of 2.988 mL. There were significant differences among the variables of the CP cannula and the other three mock cannulas concerning the drained fluid flow (P<0.01).Conclusion:
In our study, the measured mean volumes showed us that more drainage surface area provides better fluid drainage.
Texto completo:
Disponível
Coleções:
Bases de dados internacionais
Base de dados:
LILACS
Assunto principal:
Velocidade do Fluxo Sanguíneo
/
Teste de Materiais
/
Cateterismo
/
Ponte Cardiopulmonar
/
Desenho de Equipamento
/
Cânula
Tipo de estudo:
Estudo prognóstico
Idioma:
Inglês
Revista:
Rev. bras. cir. cardiovasc
Assunto da revista:
Cardiologia
/
CIRURGIA GERAL
Ano de publicação:
2018
Tipo de documento:
Artigo
País de afiliação:
Chipre
Instituição/País de afiliação:
Near East University/CY