Control de Calidad al servicio de radiología diagnóstica en los Policlínicos y Hospitales de Santiago de Cuba / Quality control to the service of diagnostic radiology in policlinics and hospitals of Santiago de Cuba
Griñan Torres, Reinaldo; Parra Caboberde, Hector; Semanat Sánchez, Lino A.
Afiliação
  • Griñan Torres, Reinaldo; Centro Provincial de Higiene y Epidemiología. Santiago de Cuba. Cuba
  • Parra Caboberde, Hector; Centro Provincial de Higiene y Epidemiología. Santiago de Cuba. Cuba
  • Semanat Sánchez, Lino A; Centro Provincial de Higiene y Epidemiología. Santiago de Cuba. Cuba
Nucleus ; (31): 43-46, 2002. tab, graf
Artigo em Espanhol | CUMED | ID: cum-41091
Biblioteca responsável: CU418.1
RESUMEN
El trabajo presenta los resultados obtenidos en el control de calidad realizado a 33 salas de rayos x diagnóstico . Se incluyen los controles al generador de rayos X , al tubo a los dispositivos de colimación y alineación, así como a los sistemas de visualización (negastoscopios), cuartos oscuros, almacenes de películas, procesadoras, cartulinas insensificadoras y chasis. Además representan los resultados del control de las dosis impartidas a los pacientes en proyecciones radiográficas estándar (tórax, columna, lumbar, abdomen, cráneo y pelvis) mediante la determinación de las dosis de entrada para un paciente de referencia. Para la evaluación de los parámetros técnicos verificados se tomó como referencia los criterios de tolerancia recomendados por la guía técnica elaborada por el centro de control Estatal de equipos Médicos de la República de Cuba. En los niveles de dosis se tomó como niveles orientativos los recomendados por las Normas básicas Internacionales de seguridad para la protección contra la radiación ionizante y para la seguridad de las fuentes de radiación(AU)
ABSTRACT
The work presents the results obtained in the quality control diagnostic x-ray equipment accomplished in 33 clinics of Santiago de Cuba City. The performed test were to the x-ray generator. X ray tube, devices collimation and alignment. As well as to the light boxes. Dark rooms, and radiographic screen-film combinations .Morover the work presents the results of the entrance doses for reference patient in radiographic projections of frequents use in clinics of Santiago de Cuba city(Chest, lumbar spine , abdomen and pelvis). For the evaluation of the verified technical parameters were used the tolerance criterions recommended by the technical guide elaborated by the State Control Center of Medical Equipment of Cuba in the case of the entrance dose were used the dose levels that recommended by the International Basic Safely Standars for protection against ionizing radiation and for safely of radiation sources(AU)
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Cuba Base de dados: CUMED Assunto principal: Controle de Qualidade / Proteção Radiológica / Monitoramento de Radiação / Controle da Radiação / Controle da Exposição à Radiação Tipo de estudo: Estudo diagnóstico / Guia de prática clínica Limite: Humanos País/Região como assunto: Cuba Idioma: Espanhol Revista: Nucleus Ano de publicação: 2002 Tipo de documento: Artigo Instituição/País de afiliação: Centro Provincial de Higiene y Epidemiología/Cuba
Buscar no Google
Coleções: Bases de dados nacionais / Cuba Base de dados: CUMED Assunto principal: Controle de Qualidade / Proteção Radiológica / Monitoramento de Radiação / Controle da Radiação / Controle da Exposição à Radiação Tipo de estudo: Estudo diagnóstico / Guia de prática clínica Limite: Humanos País/Região como assunto: Cuba Idioma: Espanhol Revista: Nucleus Ano de publicação: 2002 Tipo de documento: Artigo Instituição/País de afiliação: Centro Provincial de Higiene y Epidemiología/Cuba
...