Mínimo flujo de oxígeno necesario para soporte vital durante la simulación de reanimación post parada cardiorrespiratoria / Minimum oxygen flow needed for vital support during simulated post-cardiorespiratory arrest resuscitation
Sanz-Sanjosé, E; Ariño Irujo, JJ; Sánchez Martín, CE; González Perrino, C; López-Timoneda, F.
Afiliação
  • Sanz-Sanjosé, E; Hospital Clínico San Carlos2Servicio de Anestesiología y Reanimación. Madrid. España
  • Ariño Irujo, JJ; Al Noor Hospital. Servicio de Anestesiología y Reanimación. Abu Dabi. Emiratos ÁRabes Unidos
  • Sánchez Martín, CE; Hospital Clínico San Carlos2Servicio de Anestesiología y Reanimación. Madrid. España
  • González Perrino, C; Hospital Clínico San Carlos2Servicio de Anestesiología y Reanimación. Madrid. España
  • López-Timoneda, F; Hospital Clínico San Carlos2Servicio de Anestesiología y Reanimación. Madrid. España
Rev. esp. anestesiol. reanim ; 63(5): 261-266, mayo 2016. tab, ilus
Artigo em Espanhol | IBECS | ID: ibc-152290
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN
De acuerdo con las guías de la AHA y la ERC, la FiO2 a administrarse debería ser aquella con la que se obtuviera una SatO2 ≥ 94%. El objetivo de este estudio es determinar el mínimo flujo de oxígeno y tiempo necesarios para alcanzar una FiO2 de 0,32 y de 0,80 durante el manejo posparada cardiaca. Se emplearon diferentes reanimadores, que fueron conectados a un pulmón artificial Mark IV, SPUR II, Revivator Res-Q, O-TWO. Los flujos de oxígeno probados fueron 2, 5, 10 y 15 lpm. Los test estadísticos aplicados fueron Bonferroni y U de Mann-Whitney. Se obtuvo una FiO2 ≥ 0,32 con cualquiera de los flujos de oxígeno y reanimadores. Tras un mínimo de 75 s ventilando con 2 o 5 lpm, solo se consiguió una FiO2 de 0,80 con Mark IV. Se hallaron diferencias clínica y estadísticamente significativas (p < 0,05) con 15 lpm se necesitaron 35 s para alcanzar una FiO2 ≥ 0,80 con Mark IV (85,6 [0,3]) y Revivator (84,3 [1,5]) comparado con los 50 s que precisó SPUR II (87,1 [6,4]); con 2 lpm, todos los resucitadores alcanzaron una FiO2 ≥ 0,32 en 30 s(Mark IV (34,8 [1,3]), Revivator (35,7 [1,5]) y SPUR II (34,4 [2,1]), excepto O-TWO, que necesitó 35 s (36,3 [4,3]). Para alcanzar una FiO2 de 0,32 se podría emplear cualquiera de los resucitadores usando 2 lpm, aunque quizá el menos recomendable sería O-TWO. Si el objetivo fuera una FiO2 de 0,80, debería bastar con 10 lpm, usando preferiblemente Mark IV o Revivator Res-Q. En conclusión, atendiendo a los resultados de nuestro estudio, ante cualquier situación potencial, sería preferible emplear Revivator Res-Q o Mark IV que O-TWO o SPUR II (AU)
ABSTRACT
According to the ERC and the AHA guidelines, FiO2 should be titrated to achieve an O2Sat ≥ 94%. The aim of this study was to determine the minimum oxygen flow and time needed to reach an FiO2 of 0.32 and 0.80 during post-cardiac arrest care. An experimental analysis was performed that consisted of a simulated post-cardiac arrest situation. Different resuscitators were tested and connected to an artificial lung Mark IV, SPUR II, Revivator Res-Q, O-TWO. The oxygen flow levels tested were 2, 5, 10 and 15 lpm. Bonferroni and Mann-Whitney U tests were used. An FiO2 of 0.32 or more was obtained using any of the oxygen flow and resuscitators. Only the Mark IV achieved an FiO2 of 0.80 after a minimum of 75 s ventilating with 2 or 5 lpm. Clinical and statistical differences (P < .05) were found at 15 lpm it took 35 s to reach an FiO2 of 0.80 or more for Mark IV (85.6 [0.3]) and Revivator (84.3 [1.5]) compared to 50 s for SPUR II (87.1 [6.4]); at 2 lpm, all of the devices reached an FiO2 of ≥ 0.32 at 30 s(Mark IV (34.8 [1.3]), Revivator (35.7 [1.5]) and SPUR II (34.4 [2.1]), except for O-TWO, which took 35 s (36.3 [4.3]). Patients could be ventilated with any of the resuscitators using 2 lpm to obtain an FiO2 of 0.32, although possibly O-TWO would be the last option during the first 60 s. In order to reach an FiO2 of 0.80, ventilating with 10 lpm should be sufficient, and preferably using Mark IV or Revivator Res-Q. In conclusion, on observing the results of our study, in any possible scenario, it would be advisable to use Revivator Res-Q or Mark IV rather than O-TWO or SPUR II (AU)
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Oxigênio / Oxigenoterapia / Equipe de Assistência ao Paciente / Ventiladores Mecânicos / Reanimação Cardiopulmonar / Equipe de Respostas Rápidas de Hospitais / Parada Cardíaca Induzida Tipo de estudo: Estudo de avaliação / Guia de prática clínica Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. esp. anestesiol. reanim Ano de publicação: 2016 Tipo de documento: Artigo Instituição/País de afiliação: Al Noor Hospital/Emiratos ÁRabes Unidos / Hospital Clínico San Carlos2Servicio de Anestesiología y Reanimación/España
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Oxigênio / Oxigenoterapia / Equipe de Assistência ao Paciente / Ventiladores Mecânicos / Reanimação Cardiopulmonar / Equipe de Respostas Rápidas de Hospitais / Parada Cardíaca Induzida Tipo de estudo: Estudo de avaliação / Guia de prática clínica Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. esp. anestesiol. reanim Ano de publicação: 2016 Tipo de documento: Artigo Instituição/País de afiliação: Al Noor Hospital/Emiratos ÁRabes Unidos / Hospital Clínico San Carlos2Servicio de Anestesiología y Reanimación/España
...