Your browser doesn't support javascript.
loading
Preliminary evaluation of near-infrared vein visualization technology in the screening of palatal blood vessels
Yaprak, Emre; Kayaalti-Yuksek, Sibel.
Afiliação
  • Yaprak, Emre; Kocaeli University. Faculty of Dentistry. Department of Periodontology. İzmit. Turkey
  • Kayaalti-Yuksek, Sibel; Kocaeli University. Faculty of Dentistry. Department of Periodontology. İzmit. Turkey
Med. oral patol. oral cir. bucal (Internet) ; 23(1): e98-e104, ene. 2018. ilus, tab
Artigo em Inglês | IBECS | ID: ibc-170311
Biblioteca responsável: ES1.1
Localização: BNCS
ABSTRACT

Background:

Avoidance from palatal blood vessel rupture is a major concern during the palatal soft tissue graft surgery. There is no defined chair-side and case-specific palatal blood vessel detection approach to facilitate the harvesting process. The objective of this pilot study is to assess the feasibility of a near-infrared vein visualization system in the screening process of palatal blood vessels. Material and

Methods:

An extraoral vein visualization device (AccuVein AV400) was applied to a total of 304 hemi-maxilla of 152 individuals by two blind examiners. The study groups were classified according to their maximum inter-incisal measurements. The distances between the coronal border of the vessel image and the mid-palatal gingival margins of the adjacent teeth were measured and in each group. The correlations among the measurements were evaluated within groups.

Results:

The blood vessel to the adjacent teeth measurements exhibited no statistical difference between both examiners in all subjects (p<0.001). Correlations between the examiners gradually increased in all groups as the mouth opening rates of the subjects were increased (p<0.001).

Conclusions:

In the current state, screening of the palatal blood vessels via near-infrared vein visualization technology seems to be not suitable for every individual due to the restrictive effect of mouth opening. However, the promising results of this preliminary study demonstrated increasing consistency between the measurements of the examiners as the inter-incisal distance increase which emphasized the need an intraoral version of the device. Considering the lack of local decision-making technology for the detection of palatal blood vessels, further studies are required for development and optimization of these systems (AU)
RESUMEN
No disponible
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Palato / Veias / Vasos Sanguíneos / Administração Sanitária / Espectroscopia de Luz Próxima ao Infravermelho / Complicações Intraoperatórias Tipo de estudo: Estudo diagnóstico / Guia de prática clínica / Estudo prognóstico / Estudo de rastreamento Limite: Humanos Idioma: Inglês Revista: Med. oral patol. oral cir. bucal (Internet) Ano de publicação: 2018 Tipo de documento: Artigo Instituição/País de afiliação: Kocaeli University/Turkey

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Palato / Veias / Vasos Sanguíneos / Administração Sanitária / Espectroscopia de Luz Próxima ao Infravermelho / Complicações Intraoperatórias Tipo de estudo: Estudo diagnóstico / Guia de prática clínica / Estudo prognóstico / Estudo de rastreamento Limite: Humanos Idioma: Inglês Revista: Med. oral patol. oral cir. bucal (Internet) Ano de publicação: 2018 Tipo de documento: Artigo Instituição/País de afiliação: Kocaeli University/Turkey
...