Repeatability of the resonance frequency analysis values in implants with a new technology
Díaz-Castro, M Carmen; Falcao, Artur; López-Jarana, Paula; Falcao, Carlos; Ríos-Santos, Jose-Vicente; Fernández-Palacín, Ana; Herrero-Climent, Mariano.
Afiliação
  • Díaz-Castro, M Carmen; University of Seville. Seville. Spain
  • Falcao, Artur; Fernando Pessoa University and Porto Dental Institute. Porto. Portugal
  • López-Jarana, Paula; Fernando Pessoa University and Porto Dental Institute. Porto. Portugal
  • Falcao, Carlos; Fernando Pessoa University and Porto Dental Institute. Porto. Portugal
  • Ríos-Santos, Jose-Vicente; Dental School University of Sevilla. Seville. Spain
  • Fernández-Palacín, Ana; Dental School University of Sevilla. Seville. Spain
  • Herrero-Climent, Mariano; Fernando Pessoa University and Porto Dental Institute. Porto. Portugal
Med. oral patol. oral cir. bucal (Internet) ; 24(5): e636-e642, sept. 2019. ilus, tab
Artigo em Inglês | IBECS | ID: ibc-185682
Biblioteca responsável: ES1.1
Localização: BNCS
ABSTRACT

Background:

Assess the reliability (by means of reproducibility and repeatability) of the PenguinRFA system, analyse the ISQ values of different implant types and correlate the ISQ with the insertion torque during the placement of the implant. Material and

Methods:

120 rough surface implants were placed in bovine bone (type II and III). The implants were divided into groups, according to its design. Once the implants were in place, the exact insertion torque was registered. Then, primary stability was measured by means of the resonance frequency analysis with the PenguinRFA and the Osstell ISQ devices. In each implant two transducers of each device were used. Three measurements were obtained with each transducer.

Results:

The mean ISQ (implant stability quotient) of the whole sample is 67,70 ± 5,51. The Intraclass Correlation Coefficient (ICC) is 0,933 and 0,944 for transducers 1 and 2 respectively. The reproducibility is 0,906. The mean insertion torque is 24,54 ± 8,96N. The correlation between the ISQ and the insertion torque is 0,507 p < 0,000 (MultiPeg 1) and 0,468 p < 0,000 (MultiPeg 2) for bone type II and 0,533 p < 0,801 (MultiPeg 1) and 0,193 p < 0,140 (MultiPeg 2) for bone type III
RESUMEN
No disponible
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Implantes Dentários Limite: Animais Idioma: Inglês Revista: Med. oral patol. oral cir. bucal (Internet) Ano de publicação: 2019 Tipo de documento: Artigo Instituição/País de afiliação: Dental School University of Sevilla/Spain / Fernando Pessoa University and Porto Dental Institute/Portugal / University of Seville/Spain

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Implantes Dentários Limite: Animais Idioma: Inglês Revista: Med. oral patol. oral cir. bucal (Internet) Ano de publicação: 2019 Tipo de documento: Artigo Instituição/País de afiliação: Dental School University of Sevilla/Spain / Fernando Pessoa University and Porto Dental Institute/Portugal / University of Seville/Spain
...