Your browser doesn't support javascript.
loading
Análisis de degradación de stents metálicos en modelo de cultivo "in vitro" con células epiteliales respiratorias y fibroblastos / Degradation analysis for metal stents in an in vitro cell culture model with respiratory epithelial cells and fibroblasts
Arellano Orden, E; Sánchez López, V; Yubero Valencia, F; Márquez Martín, Eduardo; Serrano Casorrán, C; Rodríguez Panadero, F.
Afiliação
  • Arellano Orden, E; Instituto de Biomedicina de Sevilla (IbiS). Laboratorio de Enfermedades Respiratorias. España
  • Sánchez López, V; Instituto de Salud Carlos III. CIBER de Enfermedades Respiratorias (CIBERES). Madrid. España
  • Yubero Valencia, F; Universidad de Sevilla. Instituto de Ciencia de Materiales de Sevilla (ICMS). España
  • Márquez Martín, Eduardo; Instituto de Biomedicina de Sevilla (IbiS). Laboratorio de Enfermedades Respiratorias. España
  • Serrano Casorrán, C; Universidad de Zaragoza. Facultad de Veterinaria. Grupo de Técnicas Mínimamente Invasivas (GITMI). España
  • Rodríguez Panadero, F; Hospital Universitario Virgen del Rocío. Unidad Médico-Quirúrgica de Enfermedades Respiratorias (UMQER). Sevilla. España
Rev. esp. patol. torac ; 31(3): 158-173, oct. 2019. graf, tab
Artigo em Espanhol | IBECS | ID: ibc-187171
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

Objetivo:

investigar la posible degradación de stents metálicos tras co-cultivo con células respiratorias in vitro.

Método:

durante 21 días se han co-cultivado con la línea CRL-4011 (epitelial) y MRC-5 (fibroblastos) tres tipos de stents Wallstent(R) (aleación de cobalto-cromo-níquel y molibdeno), Zilver PTX(R) y Zilver Flex(R) (nitinol, aleación de níquel-titanio, con y sin liberación de paclitaxel, respectivamente). Las mismas células sin stent sirvieron como control. Los sobrenadantes de los cultivos se recogieron los días 3, 9, 15 y 21, se alicuotaron y almacenaron a -800 C. Mediante espectrometría de masas (ICP/ MS) se investigaron los niveles de titanio, cromo, níquel, cobalto y molibdeno en los sobrenadantes, y también se han analizado los niveles de esos mismos elementos en el medio de cultivo original (antes de añadirlo a los cultivos celulares).

Resultados:

en todos los experimentos se encontraron mayores niveles de elementos metálicos en los sobrenadantes recogidos en el tercer día de cultivo, tanto de células epiteliales como de fibroblastos, con diferencias estadísticamente significativas (p<0,002). Los sobrenadantes de los cultivos de células epiteliales con Wallstent mostraron los niveles más altos de níquel y cobalto respecto a los controles (p = 0,001), y los niveles de titanio fueron más altos en los cultivos de Zilver Flex y PTX, constituidos por una aleación de níquel y titanio (p <0,001).

Conclusiones:

hemos detectado una rápida liberación en el sobrenadante de todos los cultivos de los elementos constitutivos de los tres stents que incluimos en los experimentos, y con niveles muy superiores a los cultivos controles
ABSTRACT

Objective:

To investigate the possible degradation of metal stents after co-culture with respiratory cells in vitro.

Methods:

Three types of stents were co-cultured with the CRL-4011 line (epithelial) and the MRC-5 line (fibroblasts) Wallstent(R) (cobalt-chromium-nickelmolybdenum alloy), Zilver PTX(R) and Zilver Flex(R) (nitinol, nickel-titanium alloy, with and without paclitaxel release, respectively). The same stentless cells served as the control group. Culture supernatants were collected on days 3, 9, 15 and 21, aliquoted and stored at -80 ºC. The levels of titanium, chromium, nickel, cobalt and molybdenum in the supernatants were studied using inductively coupled plasma mass spectrometry (ICP-MS), and the levels of the same elements were also analyzed in the original culture medium (before adding them to the cell cultures).

Results:

In all experiments, higher levels of metal elements were found in the supernatants collected on the third day of culture, for both epithelial cells and fibroblasts, with statistically significant differences (p < 0.002). The supernatants of epithelial cell cultures with Wallstent showed the highest levels of nickel and cobalt in comparison to controls (p = 0.001), and titanium levels were higher in Zilver Flex and Zilver PTX cultures, consisting of a nickeltitanium alloy (p < 0.001).

Conclusion:

We have detected a rapid release in the supernatant of all the cultures of the constituent elements of the three stents that we included in the experiments, and with levels much higher than those of the control cultures
Assuntos
Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Técnicas In Vitro / Meios de Cultura / Células Epiteliais / Fibroblastos / Stents Metálicos Autoexpansíveis Idioma: Espanhol Revista: Rev. esp. patol. torac Ano de publicação: 2019 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario Virgen del Rocío/España / Instituto de Biomedicina de Sevilla (IbiS)/España / Instituto de Salud Carlos III/España / Universidad de Sevilla/España / Universidad de Zaragoza/España
Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Técnicas In Vitro / Meios de Cultura / Células Epiteliais / Fibroblastos / Stents Metálicos Autoexpansíveis Idioma: Espanhol Revista: Rev. esp. patol. torac Ano de publicação: 2019 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario Virgen del Rocío/España / Instituto de Biomedicina de Sevilla (IbiS)/España / Instituto de Salud Carlos III/España / Universidad de Sevilla/España / Universidad de Zaragoza/España
...