Measurement of gap between abutment and fixture in dental conical connection implants. A focused ion beam SEM observation
Carnovale, Fabio; Patini, Romeo; Peñarrocha-Oltra, David; Muzzi, Maurizio; Pistilli, Roberto; Canullo, Luigi.
Afiliação
  • Carnovale, Fabio; s.af
  • Patini, Romeo; Catholic University. Roem. Italy
  • Peñarrocha-Oltra, David; University of Valencia. Valencia. Spain
  • Muzzi, Maurizio; University of Rome 3. Rome. Italy
  • Pistilli, Roberto; s.af
  • Canullo, Luigi; s.af
Med. oral patol. oral cir. bucal (Internet) ; 25(4): e449-e454, jul. 2020. ilus, tab
Artigo em Inglês | IBECS | ID: ibc-196495
Biblioteca responsável: ES1.1
Localização: BNCS
ABSTRACT

BACKGROUND:

The aim of the authors was to examine the abutment-fixture interface in Morse-type conical implants in order to verify gaps at this level using a new microscopical approach. MATERIAL AND

METHODS:

In this in vitro study, 20 abutment-fixture complexes were prepared by sectioning (longitudinal and cross-sectional to the long axis) with a microtome and then with a focused ion beam (FIB). This is a micrometric machine tool that uses gallium ions to abrade circumscribed areas to dig deeper into the cuts obtained with the microtome in order to eliminate cut-induced artifacts. This is because the FIB abrasion is practically free from artifacts, which are normally generated by the action of the microtome blades or other techniques. Samples were then observed by scanning electron microscopy (SEM).

RESULTS:

The observation of the abraded parts with the FIB permitted measurement of the real gap between the implant-abutment components. A variable amount of gap was retrieved (from 0 to 3 μm) by the observations, confirming the non-hermetic nature of the connection. It has to be pointed out that in approximately 65% of cases, the gap accounted for less than 1 μm.

CONCLUSIONS:

The reported data confirmed that the analyzed connection system allowed for minimal gap. However, from the evidence of the present analysis, it cannot be assumed that the 2 parts of a Morse-type conical implant are fused in 1 piece, which would create a perfectly matched hermetic connection
RESUMEN
No disponible
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Implantação Dentária Endóssea / Projeto do Implante Dentário-Pivô Idioma: Inglês Revista: Med. oral patol. oral cir. bucal (Internet) Ano de publicação: 2020 Tipo de documento: Artigo Instituição/País de afiliação: Catholic University/Italy / University of Rome 3/Italy / University of Valencia/Spain

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Implantação Dentária Endóssea / Projeto do Implante Dentário-Pivô Idioma: Inglês Revista: Med. oral patol. oral cir. bucal (Internet) Ano de publicação: 2020 Tipo de documento: Artigo Instituição/País de afiliação: Catholic University/Italy / University of Rome 3/Italy / University of Valencia/Spain
...