Your browser doesn't support javascript.
loading
Influencia de las venas colaterales en la medición del flujo del acceso vascular medicante ecografía y termodilución: estudio observacional / Influence of collateral veins on flow measurement of vascular access using ultrasound and thermodilution: an observational study
Rubio-Castañeda, Francisco Javier; Sierra-Sánchez, Ana Isabel; Ferrer-López, Emilia; Fernández-Núñez, Manuel; Cantín-Lahoz, Víctor; Mateo-Sánchez, María Amaya.
Afiliação
  • Rubio-Castañeda, Francisco Javier; Hospital Universitario Miguel Servet. Zaragoza. España
  • Sierra-Sánchez, Ana Isabel; Hospital Universitario Miguel Servet. Zaragoza. España
  • Ferrer-López, Emilia; Hospital Universitario Miguel Servet. Zaragoza. España
  • Fernández-Núñez, Manuel; Hospital Universitario Miguel Servet. Zaragoza. España
  • Cantín-Lahoz, Víctor; Hospital Universitario Miguel Servet. Zaragoza. España
  • Mateo-Sánchez, María Amaya; Hospital Universitario Miguel Servet. Zaragoza. España
Enferm. nefrol ; 25(4): 330-336, octubre 2022. graf
Artigo em Espanhol | IBECS | ID: ibc-214108
Biblioteca responsável: ES1.1
Localização: ES15.1 - BNCS
RESUMEN

Introducción:

La ecografía doppler y métodos dilucionales permiten monitorizar el flujo del acceso vascular. La presencia de venas colaterales podría influir en la determinación del flujo del acceso vascular.

Objetivo:

Determinar la influencia de las venas colaterales en la medición del flujo del acceso vascular por ecografía doppler y por el método de termodilución.Material y

Método:

Estudio observacional y prospectivo. Se seleccionaron las fístulas arteriovenosas nativas que cumplían los criterios de selección. Se realizó una medición anual del flujo del acceso vascular mediante ecografía-doppler y 2 mediciones trimestrales de termodilución (termodilución-1 y termodilución-2). Para determinar la presencia de venas colaterales se empleó el ecógrafo.

Resultados:

Se analizaron 38 fístulas arteriovenosas nativas. Los hombres representaban el 78,9% de la muestra. El 23,6% presentaban venas colaterales. Analizando la totalidad de la muestra, se obtuvo correlación lineal entre los flujos por ecografía-doppler con los de termodilución-1 (0,694) y con termodilución-2 (0,678), ambas p<0,001. Al estratificar por venas colaterales, se observó correlación significativa entre ecografía-doppler con termodilución-1 (0,698) y termodilución-2 (0,696) ambas significativas (p<0,001) cuando no existían venas colaterales, correlación no significativa cuando si había venas colaterales.

Conclusiones:

Existe correlación entre la medición del flujo del acceso vascular obtenido por ecografía-doppler y termodilución; la presencia de venas colaterales modifica esta relación. Es necesario implementar programas de vigilancia del acceso vascular que incluyan diferentes métodos de monitorización, para mitigar el efecto que las venas colaterales tienen en la determinación del flujo del acceso vascular. (AU)
ABSTRACT

Introduction:

Doppler ultrasound and dilutional methods allow monitoring of vascular access flow. The presence of collateral veins may influence the determination of vascular access flow.

Objective:

To determine the influence of collateral veins on the vascular access flow measurement by Doppler ultrasound and by the thermodilution method.Material and

Method:

Observational and prospective study. Native arteriovenous fistulas that met the selection criteria were selected. An annual measurement of vascular access flow by Doppler ultrasound and two quarterly thermodilu-tion measurements (thermodilution-1 and thermodilution-2) were performed. Ultrasound was used to determine the pre-sence of collateral veins.

Results:

Thirty-eight native arteriovenous fistulae were analysed. Males accounted for 78.9% of the sample. Collateral veins were present in 23.6% of cases. Analysing the whole sample, a linear correlation was obtained between ultrasound-Doppler flows with thermodilution-1 (0.694) and thermodilution-2 (0.678), [both p<0.001]. When stratified by collateral veins, when no collateral veins were present, there was a significant correlation between Doppler ultrasound with thermodilution-1 (0.698) and thermodilution-2 (0.696) [both p<0.001)]; whereas the correlation was non-significant when collateral veins were present.

Conclusions:

There is correlation between vascular access flow measurement obtained by Doppler ultrasound and thermodilution; the presence of collateral veins modifies such relationship. It is necessary to implement vascular access surveillance programs that include different monitoring methods in order to mitigate the effect of collateral veins on vascular access flow determination. (AU)
Assuntos


Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Termodiluição / Ultrassonografia Doppler / Dispositivos de Acesso Vascular / Enfermagem em Nefrologia Limite: Humanos Idioma: Espanhol Revista: Enferm. nefrol Ano de publicação: 2022 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario Miguel Servet/España

Texto completo: Disponível Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Termodiluição / Ultrassonografia Doppler / Dispositivos de Acesso Vascular / Enfermagem em Nefrologia Limite: Humanos Idioma: Espanhol Revista: Enferm. nefrol Ano de publicação: 2022 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario Miguel Servet/España
...