Determinación del factor de calibración para el samario-153 en el activímetro de dosis CRC®-35R / Determination of dose calibrator setting of the capintec CRC®-35R for samarium-153
Plancha Mansanet, MC; Félix Fontestad, J; Caballero Calabuig, EV; Reyes Ojeda, M. D; Sopena Novales, P; Valle del Val, EM del; Abreu Sánchez, P; Cano Terol, MC; Sopena Monforte, R.
Afiliação
  • Plancha Mansanet, MC; Hospital Universitario Dr. Peset. Valencia. España
  • Félix Fontestad, J; Hospital Universitario Dr. Peset. Valencia. España
  • Caballero Calabuig, EV; Hospital Universitario Dr. Peset. Valencia. España
  • Reyes Ojeda, M. D; Hospital Universitario Dr. Peset. Valencia. España
  • Sopena Novales, P; Hospital Universitario Dr. Peset. Valencia. España
  • Valle del Val, EM del; Hospital Universitario Dr. Peset. Valencia. España
  • Abreu Sánchez, P; Hospital Universitario Dr. Peset. Valencia. España
  • Cano Terol, MC; Hospital Universitario Dr. Peset. Valencia. España
  • Sopena Monforte, R; Hospital Universitario Dr. Peset. Valencia. España
Rev. esp. med. nucl. (Ed. impr.) ; 26(6): 354-358, nov.-dic. 2007. tab, graf
Artigo em Es | IBECS | ID: ibc-69838
Biblioteca responsável: ES15.1
Localização: ES15.1 - BNCS
ABSTRACT
Determination of dose calibrator setting of the capintec CRC®-35R for samarium-153 Objetivo. Determinar el factor de calibración del activímetro para el radionucleido samario-153 (153Sm) y estudiar la influencia de la geometría en la medida de la actividad, de forma que se pueda determinar con exactitud la actividad administrada al paciente.Material y métodos. El factor de calibración del actinímetro para el 153Sm se determinó a partir del valor medio de la respuesta del detector en un canal de factor de calibración conocido, utilizando fuentes de 153Sm de uso clínico. El factor de corrección según geometría se calculó a partir del valor de actividad real indicada por el fabricante y las medidas de actividad realizadas en vial y jeringa. Resultados. El factor de calibración obtenido es de 239 ± 4. El factor de corrección que tiene en cuenta la geometría es de 0,87 ± 0,07. La actividad medida en jeringa es un 18 % superior a la medida en vial. Conclusiones. El método empleado permite medir correctamente la actividad de 153Sm, tanto en vial como en jeringa para dosificar con exactitud la actividad de Samario [153Sm] lexidronam administrada al paciente
ABSTRACT
Objetive. The aim of this work is to determine the calibration factor for Sm-153 and evaluate the influenceof the geometry in the activity measurements in order to administer accurately the activity to the patient.Methods. The calibration factor for Sm-153 was determinated using the Sm-153 sources commonly used in clinical practice and the calibrator response in a known calibration setting. The geometry correction factor for the vial and the plastic syringe was calculated using the real activity indicated by the manufacturer and the activity measured in the vial and in the plastic syringe.Results. The calibration factor obtained is 239 ± 4 and the correction factor which takes in consideration the geometry is 0,87 ± 0,07. The activity measured in the syringe is eighteen percent higher to activity measured in the vial. Conclusion. This method allows to measure accurately the activity of Sm-153 in vial and syringe applying this geometry correction factor to determine accurately the activity administered to the patient (AU)
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Compostos Organometálicos / Compostos Organofosforados / Radioisótopos / Medicina Nuclear Idioma: Espanhol Revista: Rev. esp. med. nucl. (Ed. impr.) Ano de publicação: 2007 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario Dr. Peset/España
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Compostos Organometálicos / Compostos Organofosforados / Radioisótopos / Medicina Nuclear Idioma: Espanhol Revista: Rev. esp. med. nucl. (Ed. impr.) Ano de publicação: 2007 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Universitario Dr. Peset/España
...