Your browser doesn't support javascript.
loading
Evaluación de un método POCT (pruebas a la cabecera del paciente) para la determinación de hemoglobina glucosilada / Evaluation of a POCT (point of care testing) assay for haemoglobin A1c
Fernández Ramos, Ana María; Cobos Muñoz, Gemma; Aguilera Castillo, Alejandra; Enguíx Armada, Alfredo.
Afiliação
  • Fernández Ramos, Ana María; Hospital Virgen de la Victoria. Málaga. España
  • Cobos Muñoz, Gemma; Hospital Virgen de la Victoria. Málaga. España
  • Aguilera Castillo, Alejandra; Hospital Virgen de la Victoria. Málaga. España
  • Enguíx Armada, Alfredo; Hospital Virgen de la Victoria. Málaga. España
Rev. lab. clín ; 2(3): 148-152, jul.-sept. 2009. tab, ilus
Artigo em Espanhol | IBECS | ID: ibc-85180
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN
Introducción. La hemoglobina glucosilada (HbA1C) presenta un papel fundamental en el control y prevención de complicaciones en el paciente diabético, puesto que permite la monitorización de sus valores de glucemia en los 3 meses previos al análisis. Introducción. Debido a la importancia de su determinación han aparecido diferentes métodos de cuantificación. El objeto de este trabajo es evaluar el método POCT (point of care testing ‘pruebas a la cabecera del paciente’) mediante el aparato In2it de Bio-Rad. Material y métodos. Se realizó una comparación entre 2 métodos de cuantificación el HPLC (high–performance liquid chromatography ‘cromatografía líquida de alta resolución o de alta presión’) (ADAMS-A1c HA-8160, Menarini Diagnostics, Florencia, Italia) y el POCT (In2it, Bio-Rad, California, EE.UU.). Para esto, se compararon los resultados de 60 muestras, además de determinar la imprecisión intraserial y la imprecisión interdiaria de ambos aparatos. Resultados. La comparación entre ambos métodos mostró una buena correlación entre el HPLC y el POCT Y=0,4000+1,000 X. Los intervalos de confianza del 95% para la coordenada en el origen y la pendiente fueron de −0,1500 a 0,6500 y de 0,9565 a 1,0769, respectivamente. Resultados. La imprecisión intraserial se situó entre el 2,5 y el 3% (del 2,20 al 3,04%) y la imprecisión interdiaria se situó entre el 3,53 y el 2,95% para valores altos y bajos, respectivamente. Discusión. El método POCT ha demostrado cumplir todos los requisitos técnicos necesarios para su validación como método para la determinación de la HbA1C, ya que presenta igualdad de prestaciones analíticas con el método HPLC. Además, es apropiado para el seguimiento de pacientes diabéticos en atención primaria, dada su facilidad y sencillez de manejo (AU)
ABSTRACT
Introduction. The HbA1c is related to the control and prevention of complications in diabetics. This parameter enables glucose control to be monitored in the three months previous to the determination. Introduction. Due to the significance of its determination, different measurement methods have been developed. The aim of this work is to evaluate the POCT (point of care testing) method used by the In2it BioRad. Material and methods. We report on a comparison between two

methods:

a HPLC (high–performance liquid chromatography) (ADAMS-A1c HA-8160, Menarini) and the POCT test (In2it, BioRad). To determine the within and between day imprecision, 60 samples were analysed and compared. Results. Both methods showed good correlations (Y=0.4000+1.000 X). The 95% confidence intervals for the coordinate at the origin and the slope were (−0.1500 to 0.6500) and (0.9565 to 1.0769), respectively. Within day imprecision was 2.5–3% (2.20–3.04%) and between day imprecision was 3.53% and 2.95% for high and low values, respectively. Discussion. The POCT test (BioRad System) has demonstrated that it fulfils all the technical requirements needed for its validation as an HbA1c analysis method as it has a similar performance to the HPLC method The POCT method has shown to be suitable for the follow-up of diabetic patients in primary care due its ease and simplicity of use (AU)
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Hemoglobinas Glicadas / Sistemas Automatizados de Assistência Junto ao Leito / Indicadores e Reagentes Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. lab. clín Ano de publicação: 2009 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Virgen de la Victoria/España
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Hemoglobinas Glicadas / Sistemas Automatizados de Assistência Junto ao Leito / Indicadores e Reagentes Limite: Feminino / Humanos / Masculino Idioma: Espanhol Revista: Rev. lab. clín Ano de publicação: 2009 Tipo de documento: Artigo Instituição/País de afiliação: Hospital Virgen de la Victoria/España
...