Your browser doesn't support javascript.
loading
Freehand-SPECT con 99mTc-HDP como herramienta para guiar la biopsia percutánea de lesiones esqueléticas detectadas en la gammagrafía ósea / Freehand-SPECT with 99mTc-HDP as tool to guide percutaneous biopsy of skeletal lesions detected on bone scintigraphy
Rietbergen, DDD; Meershoek, P; Oosterom, MN van; Roestenberg, M; Erkel, AR van; Smit, F; der Hage, JA van; Valdés Olmos, RA; Leeuwen, FWB van.
Afiliação
  • Rietbergen, DDD; Leiden University Hospital. Department of Radiology and Nuclear Medicine. Países Bajos
  • Meershoek, P; Leiden University Medical Center. Department of Radiology. Interventional Molecular Imaging Laboratory. Leiden. Países Bajos
  • Oosterom, MN van; Leiden University Medical Center. Department of Radiology. Interventional Molecular Imaging Laboratory. Leiden. Países Bajos
  • Roestenberg, M; Leiden University Medical Centre. Department of Infectious Diseases. Leiden. Países Bajos
  • Erkel, AR van; Leiden University Hospital. Department of Radiology and Nuclear Medicine. Países Bajos
  • Smit, F; Leiden University Hospital. Department of Radiology and Nuclear Medicine. Países Bajos
  • der Hage, JA van; Leiden University Hospital. Department of Surgery. Países Bajos
  • Valdés Olmos, RA; Leiden University Hospital. Department of Radiology and Nuclear Medicine. Países Bajos
  • Leeuwen, FWB van; Leiden University Medical Center. Department of Radiology. Interventional Molecular Imaging Laboratory. Leiden. Países Bajos
Rev. esp. med. nucl. imagen mol. (Ed. impr.) ; 38(4): 218-223, jul.-ago. 2019. ilus, tab, graf
Artigo em Espanhol | IBECS | ID: ibc-188692
Biblioteca responsável: ES1.1
Localização: BNCS
RESUMEN

OBJETIVOS:

Determinar la factibilidad de la freehand-SPECT en la identificación de lesiones óseas con captación de 99mTc-HDP evaluando los datos generados mediante la utilización de sistemas de navegación con realidad aumentada y virtual. MATERIAL Y

MÉTODOS:

Se generaron 21 imágenes 3D utilizando freehand-SPECT con una gammacámara manual en 20 pacientes referidos para gammagrafía ósea con 99mTc-HDP. Las imágenes freehand-SPECT y las gammagrafías fueron comparadas y se analizó el grado de concordancia. Además, se evaluó la factibilidad de navegación hacia las lesiones óseas hipercaptantes.

RESULTADOS:

En el 86% de los casos freehand-SPECT mostró una buena concordancia con las imágenes correspondientes de la gammagrafía ósea. En lesiones con una señal lesión/fondo de>1,36 freehand-SPECT pudo automáticamente proporcionar puntos de referencia segmentados con finalidad de navegación. En el 14% de los casos (índice lesión/fondo valor promedio 1,82, rango 1,0-3,4) las imágenes freehand-SPECT mostraron concordancia intermedia debido a que estaban localizadas en regiones anatómicas difíciles o asociadas con una gammagrafía ósea negativa y fueron consideradas como no apropiadas para la navegación dirigida.

CONCLUSIÓN:

En este estudio piloto, se encontró un 86% de los casos apropiados para propósitos de navegación con una buena concordancia entre freehand-SPECT y la gammagrafía ósea. Un índice lesión/fondo de 1,36 o más facilitó la navegación con freehand-SPECT. La alta calidad de las imágenes generadas con freehand-SPECT potencialmente asegura una exitosa estrategia de navegación para biopsias óseas guiadas
ABSTRACT
PURPUSE To assess the feasibility of using freehand Single Photon Emission Computed Tomography (freehandSPECT) for the identification of technetium-99m-hydroxydiphosphonate (99mTc-HDP) positive bone lesions and to evaluate the possibility of using these imaging data-sets for augmented- and virtual-reality based navigation approaches. MATERIAL AND

METHODS:

In 20 consecutive patients referred for scintigraphy with 99mTc-HDP, 21 three-dimensional freehandSPECT-images were generated using a handheld gamma camera. Concordance of the two different data sets was ranked. Furthermore, feasibility of segmenting the hotspot of tracer accumulation for navigation purposes was assessed.

RESULTS:

In 86% of the cases freehandSPECT images showed good concordance with the corresponding part of the scintigraphic images. In lesions with a signal to background ratio (SBR) >1.36, freehandSPECT provided an automatically segmented reference point for navigation purposes. In 14% of the cases (average SBR 1.82, range 1.0-3.4) freehandSPECT images showed intermediate concordance due to difficult anatomical area or negative bone scintigraphy and could not be used as navigation targets.

CONCLUSION:

In this pilot study, in 86% of the cases freehandSPECT demonstrated good concordance with traditional scintigraphy. A lesion with a SBR of 1.36 or more was suitable for navigation. These high-quality freehandSPECT images supported the future exploration navigation strategies, e. g. guided needle biopsies
Assuntos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Biópsia por Agulha / Doenças Ósseas / Câmaras gama / Tomografia Computadorizada de Emissão de Fóton Único / Medronato de Tecnécio Tc 99m / Biópsia Guiada por Imagem Limite: Humanos Idioma: Espanhol Revista: Rev. esp. med. nucl. imagen mol. (Ed. impr.) Ano de publicação: 2019 Tipo de documento: Artigo Instituição/País de afiliação: Leiden University Hospital/Países Bajos / Leiden University Medical Center/Países Bajos / Leiden University Medical Centre/Países Bajos
Buscar no Google
Coleções: Bases de dados nacionais / Espanha Base de dados: IBECS Assunto principal: Biópsia por Agulha / Doenças Ósseas / Câmaras gama / Tomografia Computadorizada de Emissão de Fóton Único / Medronato de Tecnécio Tc 99m / Biópsia Guiada por Imagem Limite: Humanos Idioma: Espanhol Revista: Rev. esp. med. nucl. imagen mol. (Ed. impr.) Ano de publicação: 2019 Tipo de documento: Artigo Instituição/País de afiliação: Leiden University Hospital/Países Bajos / Leiden University Medical Center/Países Bajos / Leiden University Medical Centre/Países Bajos
...