RESUMO

OBJETIVO:

Determinar a reprodutibilidade da espessura da camada de fibras nervosas da retina peripapilar e dos parâmetros da cabeça do nervo óptico em olhos normais, utilizando a tomografia de coerência óptica de domínio espectral (SD-OCT).

MÉTODOS:

Trinta e quatro olhos de 34 indivíduos saudáveis foram incluídos no estudo. O protocolo do cubo do disco óptico 200x200 do OCT Cirrus foi utilizado para gerar 3 imagens de cada olho para avaliar a reprodutibilidade. O coeficiente de correlação intraclasse (ICC) foi calculado para as medidas da espessura da camada de fibras nervosas da retina e da cabeça do nervo óptico.

RESULTADOS:

A correlação foi excelente para todas as medidas da espessura da camada de fibras nervosas da retina (ICC entre 0,87 e 0,98). O mesmo ocorreu com os parâmetros da cabeça do nervo óptico que, com exceção da razão E/D vertical (ICC 0,56), apresentou ICC entre 0,83 e 0,99.

CONCLUSÃO:

As medidas de espessura da camada de fibras nervosas da retina peripapilar e os parâmetros da cabeça no nervo óptico demonstraram uma excelente reprodutibilidade com o OCT Cirrus, indicando que este aparelho poderá ser uma ferramenta útil no estudo do glaucoma. Este estudo apresenta algumas limitações, como o pequeno número de casos avaliados, sendo necessários maiores estudos para corroborar nossos achados.
ABSTRACT

PURPOSE:

To determine the reproducibility of the peripapillary retinal nerve fiber layer thickness and optic nerve head parameters measured with the optical coherence tomography (SD-OCT) in normal eyes.

METHODS:

Thirty-four eyes of 34 healthy patients were included in the study. The Cirrus OCT optic disc cube 200x200 protocol was used to obtain three scans from each eye to evaluate the reproducibility. Intraclass correlation coefficient (ICC) was calculated for the measurements of retinal nerve fiber layer thickness and optic nerve head parameters.

RESULTS:

The correlation was excellent for all measures of retinal nerve fiber layer thickness (ICC between 0.87 and 0.98). The same happened with the optic nerve head parameters (ranging from 0.83 to 0.99), with the exception of E/D vertical ratio (0.56).

CONCLUSION:

Measurements of peripappilary retinal nerve fiber layer thickness and optic nerve head parameters showed excellent reproducibility with Cirrus OCT, indicating that this instrument may be useful in the study of glaucoma. This study has some limitations, including the small number of cases studied, further studies are needed to corroborate our findings.
Assuntos


Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Disco Óptico / Nervo Óptico / Retina / Glaucoma / Reprodutibilidade dos Testes / Tomografia de Coerência Óptica / Fibras Nervosas Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Rev. bras. oftalmol Assunto da revista: Oftalmologia Ano de publicação: 2012 Tipo de documento: Artigo

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Disco Óptico / Nervo Óptico / Retina / Glaucoma / Reprodutibilidade dos Testes / Tomografia de Coerência Óptica / Fibras Nervosas Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Rev. bras. oftalmol Assunto da revista: Oftalmologia Ano de publicação: 2012 Tipo de documento: Artigo
...