Uso seletivo dos músculos da mão e antebraço durante a inserção de mini-implantes: um torquímetro natural / Selective use of hand and forearm muscles during mini-implant insertion: a natural torque meter
Estelita, Sérgio; Janson, Guilherme; Chiqueto, Kelly; Ferreira, Eduardo Silveira.
Afiliação
  • Estelita, Sérgio; Universidade Federal do Rio Grande do Sul. Faculdade de Odontologia. Porto Alegre. BR
  • Janson, Guilherme; Universidade de São Paulo. Faculdade de Odontologia. Bauru. BR
  • Chiqueto, Kelly; Universidade de São Paulo. Faculdade de Odontologia. Bauru. BR
  • Ferreira, Eduardo Silveira; Universidade Federal do Rio Grande do Sul. Faculdade de Odontologia. Porto Alegre. BR
Ortodontia ; 45(5): 565-573, set.-out. 2012. ilus, tab, graf
Artigo em Português | LILACS, BBO - Odontologia | ID: lil-714058
Biblioteca responsável: BR22.1
RESUMO
Proposição comparar o torque máximo produzido pelaação de dois diferentes grupos musculares e sua influência na limitaçãodo toque de inserção e prevenção à fratura dos mini-implantes. Materiale

Métodos:

uma amostra de 87 profissionais foi avaliada quanto àintensidade máxima do torque aplicado à chave de inserção pela açãoconjunta dos dedos polegar e indicador (torque digital máximo - TOM),e pela ação supinadora do antebraço (torque braquial máximo - TBM).Foram fraturados 90 mini-implantes igualmente distribuídos entrenove diferentes diâmetros, determinando o torque de fratura (TF). Oíndice de resistência à fratura (lRF)foi calculado a partir das relaçõesIRF_TOM= TF/TDM e IRF_TBM = TF/TBM. As variáveis foram comparadaspelos testes te Anova, utilizando-se um nível de significânciade 5%.

Resultados:

o valor de TDM foi significantemente menor queTBM, e ambos foram menores no sexo feminino. TF mostrou aumentosignificante para cada 0,1 mm de aumento do diâmetro. IRF_TOMfoi maior que IRF_TBM para todos os diâmetros de mini-implantes.Entretanto, IRF_TOM > 1 foi observado apenas para mini-implantescom diâmetro maior que 1,5 mm, enquanto IRF_TBM > 1 ocorreu apartir do diâmetro de 1,7 mm para o sexo feminino e 1,8 mm para omasculino.

Conclusão:

o valor médio do torque digital foi 42% menordo que o torque braquial e mostrou-se mecanicamente mais seguro ebiologicamente mais compatível, prevenindo a fratura de mini-implantescom diâmetro maior ou igual a 1,5 mm devido à limitação do torquemédio de inserção em 15 Nem.
ABSTRACT

Objective:

to compare the maximum torque producedby different muscular groups and its influence on mini-implant insertiontorque and fracture prevention. Materiais and

Methods:

eighty-sevenprofessionals were evaluated regarding the maximum torque producedon a screwdriver by a combined action between the thumb and indexfingers (Maximum Digital Torque - MDT), and by the forearm supinationmovement (Maximum Brachial Torque - MBT). Ninety mini-implantsdistributed in nine different diameters were fractured to determine thefracture torque (FT). The fracture resistance index (FRI) was obtainedfrom FRI_MDT = FT/MDT and FRI_MBT = FT/MBT. Anova and t testswere used to compare the groups at a significance levei of 5%.

Results:

the MOT was smaller than MBr, and both were smaller in female. TheFT increased for each 0.1 mm in diameter incremento The FRI_MDT wasgreater than FRI_MBT for ali the diameters. The FRI_MDT> 1was foundwhen the diameter was greater than or equal to 1.5 mm. The FRLMBT> 1occurred with diameters equal or greater than 1.7mm for femalesand 1.8 mm for males.

Conclusion:

the digital torque was 42% smallerthan brachial torque, and it was mechanically safer and biologicallymore compatible, allowing fracture prevention of 1.5 mm or thickermini-implant diameter due to insertion torque limitation at 15 Ncm.
Assuntos
Buscar no Google
Coleções: Bases de dados internacionais Base de dados: BBO - Odontologia / LILACS Assunto principal: Gestão de Riscos / Parafusos Ósseos / Torque / Força Muscular Tipo de estudo: Estudo de etiologia Idioma: Português Revista: Ortodontia Assunto da revista: Odontologia Ano de publicação: 2012 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade Federal do Rio Grande do Sul/BR / Universidade de São Paulo/BR
Buscar no Google
Coleções: Bases de dados internacionais Base de dados: BBO - Odontologia / LILACS Assunto principal: Gestão de Riscos / Parafusos Ósseos / Torque / Força Muscular Tipo de estudo: Estudo de etiologia Idioma: Português Revista: Ortodontia Assunto da revista: Odontologia Ano de publicação: 2012 Tipo de documento: Artigo País de afiliação: Brasil Instituição/País de afiliação: Universidade Federal do Rio Grande do Sul/BR / Universidade de São Paulo/BR
...