Resposta vascular após implante de stents liberadores de biolimus A9 compolímero bioabsorvível e stents liberadores de everolimus com polímero durável. Resultados da análise de tomografia de coerência óptica do estudo randomizado BIOACTIVE / Vascular response after implantation of biolimus A9-eluting stent with bioabsorbable polymer and everolimus-eluting stents with durable polymer. Results of the optical coherence tomography analysis of the BIOACTIVE randomized trial
Chamié, Daniel; Almeida, Breno O; Grandi, Fábio; M. Filho, Evandro; Costa Junior, J. Ribamar; Costa, Ricardo; Staico, Rodolfo; Siqueira, Dimytri; Tanajura, Luiz Fernando; Centemero, Marinella; Chaves, Área J; Abizaid, Andrea; Sousa, Amanda G. M. R; Abizaid, Alexandre.
Afiliação
  • Chamié, Daniel; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Almeida, Breno O; Hospital Santa Marcelina. São Paulo. BR
  • Grandi, Fábio; Cardiovascular Research Center. São Paulo. BR
  • M. Filho, Evandro; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Costa Junior, J. Ribamar; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Costa, Ricardo; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Staico, Rodolfo; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Siqueira, Dimytri; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Tanajura, Luiz Fernando; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Centemero, Marinella; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Chaves, Área J; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Abizaid, Andrea; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Sousa, Amanda G. M. R; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Abizaid, Alexandre; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
Rev. bras. cardiol. invasiva ; 23(1): 28-37, abr.-jun.2015. ilus, tab
Artigo em Português | LILACS, Sec. Est. Saúde SP, SESSP-IDPCPROD, Sec. Est. Saúde SP | ID: lil-782172
Biblioteca responsável: BR44.1
RESUMO
No estudo BIOACTIVE, avaliamos as respostas vasculares após implante do stent eluidor de biolimus A9 (SEB; BioMatrix®) e o stent eluidor de everolimus (SEE; XIENCE V®). Apresentamos a análise detomografia de coerência óptica (OCT) 6 meses pós-intervenção.

Métodos:

Os pacientes foram randomizados para tratamento com SEB (n = 22) ou SEE (n = 18). O desfecho primário foi a frequência de hastes não cobertas e mal apostas pela OCT.

Resultados:

A OCT foi realizada em 26 pacientes (SEB n = 15; SEE n = 11) e foram analisadas 749 imagens tomográficas e 7.725 hastes de stent. SEB e SEE apresentaram áreas luminais e dos stents semelhantes.A área de hiperplasia neointimal, a espessura neointimal e o porcentual de obstrução intra-stent (8,44 ± 5,10% vs. 9,21 ± 6,36%; p = 0,74) foram similares. As taxas de hastes não cobertas (SEB 2,10 ± 3,60% vs. SEE 2,46 ± 2,15%; p = 0,77) e mal apostas (SEB 0,48 ± 1,48% vs. SEE 0,44 ± 1,05%; p = 0,94) foram baixas e semelhantes. A frequência de frames com sinais compatíveis com infiltrado inflamatório peri-haste foi baixa e similar entre SEB (15,53 ± 20,77%) e SEE (11,70 ± 27,51%; p = 0,68).

Conclusões:

Stents farmacológicos de segunda geração SEB e SEE se mostraram igualmente eficientes emsuprimir a formação neointimal aos 6 meses, com respostas vasculares favoráveis. A frequência de framescom sinais de infiltrado peri-haste por paciente foi baixa, e menor do que a historicamente observada comos stents farmacológicos de primeira geração...
ABSTRACT
In BIOACTIVE study, we evaluated vascular responses after the implant of biolimus A9-eluting stent (BES; BioMatrixTM) and the everolimus-eluting stent (EES; XIENCE VTM). In this study, we present the optical coherence tomography analysis (OCT) 6 months post-intervention.

Methods:

Patients were randomized to treatment with BES (n = 22) or EES (n = 18). The primary outcome was the frequency of non-covered, poorly positioned struts by OCT.

Results:

OCT was performed in 26 patients (BES n = 15; EES n = 11) and 749 tomographic images and 7,725stent struts were analyzed. BES and EES showed similar luminal and stent areas. Neointimal hyperplasia area, neointimal thickness and the percentage of in-stent obstruction (8.44 ± 5.10% vs. 9.21 ± 6.36%; p = 0.74) were similar. The rates of not covered struts (BES 2.10 ± 3.60% vs. ESS 2.46 ± 2.15%, p = 0.77) and poorly positioned struts (BES 0.48 ± 1.48% vs. EES 0.44 ± 1.05%, p = 0.94) were similarly low. The frequency of frames with signs consistent with peri-strut inflammatory infiltrate was low and similar between BES (15.53 ± 20.77%) and EES (11.70 ± 27.51%; p = 0.68).

Conclusions:

The second-generation drug-eluting stents BES and EES were equally effective at suppressing the neointimal formation after 6 months, with favorable vascular responses. The frequency of frameswith peri-strut infiltrate signals per patient was low, and lower than that observed historically with firstgenerationdrug-eluting stents...
Assuntos

Texto completo: Disponível Coleções: Bases de dados nacionais / Brasil Base de dados: LILACS / Sec. Est. Saúde SP / SESSP-IDPCPROD Assunto principal: Doença das Coronárias / Tomografia de Coerência Óptica / Stents Farmacológicos Tipo de estudo: Ensaio clínico controlado / Estudo diagnóstico / Estudo observacional / Fatores de risco Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Rev. bras. cardiol. invasiva Ano de publicação: 2015 Tipo de documento: Artigo Instituição/País de afiliação: Cardiovascular Research Center/BR / Hospital Santa Marcelina/BR / Instituto Dante Pazzanese de Cardiologia/BR

Texto completo: Disponível Coleções: Bases de dados nacionais / Brasil Base de dados: LILACS / Sec. Est. Saúde SP / SESSP-IDPCPROD Assunto principal: Doença das Coronárias / Tomografia de Coerência Óptica / Stents Farmacológicos Tipo de estudo: Ensaio clínico controlado / Estudo diagnóstico / Estudo observacional / Fatores de risco Limite: Feminino / Humanos / Masculino Idioma: Português Revista: Rev. bras. cardiol. invasiva Ano de publicação: 2015 Tipo de documento: Artigo Instituição/País de afiliação: Cardiovascular Research Center/BR / Hospital Santa Marcelina/BR / Instituto Dante Pazzanese de Cardiologia/BR
...