Desarrollo de test 3D automatizado en la Fundación Boliviana de Oftalmología durante la gestión 2014, para valorar estereopsis / Development of 3D automated test in the Bolivian Ophthalmology Foundation during 2014 management, to assess stereopsis
Céspedes Oporto, Víctor Hugo; Ortuño Lazarte, Patricia Elizabeth.
Afiliação
  • Céspedes Oporto, Víctor Hugo; Fundación Boliviana de Oftalmología. BO
  • Ortuño Lazarte, Patricia Elizabeth; Fundación Boliviana de Oftalmología. BO
Gac. méd. boliv ; 39(1): 16-19, jun. 2016. ilus, graf, mapas, tab
Artigo em Espanhol | LILACS | ID: lil-797287
Biblioteca responsável: BO138.1
RESUMEN

Objetivos:

desarrollar un test 3D automatizado para valorar la estereopsis, test que podría ser ampliamente difundido con el actual auge de la tecnología 3D en video juegos.

Métodos:

se realizó un estudio comparativo, prospectivo, transversal, inicialmente se desarrolló el test 3D con Scratch versión 1.4 y posteriormente se realizó el mismo en 19 personas entre 15 y 40 años con agudeza visual de 20/20 ó 20/25, se compararon los resultados con los obtenidos con el Test de Titmus.

Resultados:

encontramos una Agudeza Visual Estereoscópica (AVE) media de 55,32 arcos por segundo con el test 3D automatizado, con el Test de Titmus encontramos una AVE media de 57,89. Realizamos el análisis estadístico con el programa SPSS, con un Alfa de Cronbach de 0,988 lo que indica que existe correlación entre los resultados obtenidos por ambos test.

Conclusiones:

consideramos que el test 3D automatizado es válido para su utilización en posteriores investigaciones y se constituye en una buena opción para valoración de estereopsis.
ABSTRACT

Objective:

to develop a 3D automated test to assess stereopsis, test could be widely disseminated with the current boom of 3D technology in video games.

Methods:

a comparative, prospective, cross-sectional study, initially the 3D test was developed with Scratch version 1.4 and later it was performed in 19 people between 15 and 40 years with visual acuity of 20/20 or 20/25, were compared the results with those obtained with the Titmus test.

Results:

we found an average Stereoscopic Visual Acuity (AVE) of 55.32 arcs per second with automated 3D test with the Titmus test found a mean of 57.89 AVE. We performed statistical analysis using the SPSS program, with a Cronbach’s alpha of 0.988 indicating that there is a correlation between the results obtained by both tests.

Conclusion:

we believe that 3D automated test is valid for use in further research and constitutes a good choice for assessment of stereopsis.
Assuntos

Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Percepção de Profundidade Tipo de estudo: Estudo observacional Limite: Adolescente / Adulto / Feminino / Humanos / Masculino País/Região como assunto: América do Sul / Bolívia Idioma: Espanhol Revista: Gac. méd. boliv Assunto da revista: Medicina Ano de publicação: 2016 Tipo de documento: Artigo País de afiliação: Bolívia Instituição/País de afiliação: Fundación Boliviana de Oftalmología/BO
Texto completo: Disponível Coleções: Bases de dados internacionais Base de dados: LILACS Assunto principal: Percepção de Profundidade Tipo de estudo: Estudo observacional Limite: Adolescente / Adulto / Feminino / Humanos / Masculino País/Região como assunto: América do Sul / Bolívia Idioma: Espanhol Revista: Gac. méd. boliv Assunto da revista: Medicina Ano de publicação: 2016 Tipo de documento: Artigo País de afiliação: Bolívia Instituição/País de afiliação: Fundación Boliviana de Oftalmología/BO
...